Operating Manual V8.2/03.2017
Non Invasive Road Sensor
G. Lufft Mess- und Regeltechnik GmbH, Fellbach, Germany
55
19.5.13 Message: Station Reset with Reset of Service Level and Lamp Operating
Hours, Renew Adjustment
This command shall only be executed after replacement of the lamp unit. It clears service
level and lamp operating hours, then initiates a station reset. After the reset a new automatic
adjustment will be executed.
Important note:
For the automatic adjustment, the same conditions that are required for
commissioning (see page 24) must be met and the instructions for commissioning must be
observed, otherwise the new adjustment of the sensor cannot be completed successfully.
Command:
aXRs!
Response:
aXRok<CR><LF>
The response is followed by the station reset, i.e. the station will be offline for a few seconds.
Example:
0XRs!
0XRok<CR><LF>
19.5.14 Message: Station Reset with New Adjustment
This command initiates a station reset. After the reset a new automatic adjustment will be
executed.
Important note:
For the automatic adjustment, the same conditions that are required for
commissioning (see page 24) must be met and the instructions for commissioning must be
observed, otherwise the new adjustment of the sensor cannot be completed successfully.
Command:
aXRa!
Response:
aXRok<CR><LF>
The response is followed by the station reset, i.e. the station will be offline for a few seconds.
Example:
0XRa!
0XRok<CR><LF>