7-2
7.4 TEMPERATURE FAILURE.
If the temperature reading on the display begins to flash, then the measured
temperature is beyond the operating temperature range of the instrument. To restore proper operation, reduce or
increase the temperature as necessary. If this message is seen in error, then it is possible that the temperature sensor,
located on the electrometer board, has failed and should be replaced by a technician.
7.5 LOW POWER.
If the message "Low Batt" is shown on the display, then it is likely that the primary batteries
need to be replaced or recharged. Ensure that when replacing batteries (Section 6.2), the battery selection switch
(Figure 6-3) is in the correct position for rechargeable or non-rechargeable batteries. Alternatively, operation is
possible if the 115 VAC power cord is properly connected into a source of 115 VAC power. If operation is not
restored by this action, also ensure that the circuit breaker on the back of the instrument is not toggled to the OFF
(down) position.
7.6 LEAKS.
Perform the can assembly and the air circuit leak tests if a leak is suspected. The leak test procedure
for the Model 3100 ensures that the air circuit is functioning properly and water ingress will be limited to ensure
compliance with MIL-STD-810G Method 506 Procedures I and III.
7.6.1 Can Assembly Leak Test.
a.
Open the front panel of the Model 3100 by loosening the four latches.
b.
Loosen an air hose fitting from the desiccant switch. See Figure 7-1.
Figure 7-1. Loosening Air Hose.
c.
Apply grease to the front-panel gasket as required.
d.
Close the front panel by securing all four latches tightly.
e.
Attach the handheld vacuum pump to the OUT port located under the air filter. See Figure 7-2.
f.
Plug the IN port. See Figure 7-2.
Figure 7-2. Vacuum.
g.
Apply a vacuum of -5 inches Hg.
h.
If the vacuum is held, proceed to Air Circuit Leak Test, Section 7.6.3.
Содержание 3100
Страница 1: ...LUDLUM MODEL 3100 TRITIUM IN AIR MONITOR July 2019 ...
Страница 2: ...LUDLUM MODEL 3100 TRITIUM IN AIR MONITOR July 2019 ...
Страница 11: ...1 2 Figure 1 2 Front View Model 3100 ...
Страница 31: ......
Страница 34: ......
Страница 35: ...5 3 Figure 5 5 Rotary Draw Latches Detail View BF Figure 5 6 Rack Mount Safety Strap ...
Страница 36: ......