6-1
CHAPTER 6
SCHEDULED MAINTENANCE
6.1 CLEANING.
Cleaning the Model 3100 is important in keeping it in working order and in helping the
instrument last. The following is the procedure for cleaning the instrument.
6.1.1 Preparation.
Shut the instrument down properly before cleaning.
a.
Turn the MODE SWITCH to OFF.
b. Disconnect the AC power cord if attached.
6.1.2 Clean With Care.
CAUTION
Do not immerse the instrument in any liquid.
Gently wipe all exposed surfaces with a damp cloth, using only water as a wetting agent. Allow instrument to dry
thoroughly before reuse.
6.2 BATTERY REPLACEMENT.
Turn the MODE switch to OFF, and disconnect the AC power cord if attached
before replacing the batteries.
6.2.1 Remove Batteries.
Open the battery lid by removing the screws (Figure 6-1). Remove the 8 NIMH AA
batteries from the compartment (see Figure 6-2).
Figure 6-1. Battery Lid Removal