1-5
Table 1-1. Reference Data, continued.
Gamma
Compensation
Allows for tritium monitoring
in up to 5 mR/hr (0.05
mSv/h) gamma fields
Display Range
0-20,000 µCi/m
3
(0 to 740
MBq/m
3
)
Battery
Dependence
Calibration changes <5%
within battery check limits
Temperature
32 to 122
o
F (0 to 50
o
C)
Humidity Range
40 - 90% RH maintaining
response ±15% of reading
at 40% RH and 71 °F (22
°C)
Effective Range of
Measurement
From 1 to 10,000 µCi/m
3
(0
to 370 MBq/m
3
)
Minimum Detectable
Activity (MDA)
2 µCi/m
3
(0.074 MBq/m
3
)
Mode Switch
Positions
Off, Check, Measure,
Sample (Pump On), and
Purge
Response Time
Approximately 60 seconds
Construction
Rugged, gasketed,
waterproof
(1)
aluminum
enclosure with external
user-replaceable 0.2 µm
PTFE particulate filter
Linearity
Reading within 20% of true
value
LCD Display
2.7 in. (6.9 cm) diagonal
transflective backlit LCD
housed inside the
electronics enclosure
Pollution
Degree 3 (As defined by
IEC 644, “Occurs when
conductive pollution or dry
nonconductive pollution
becomes conductive due
to condensation.”)
Audio
75 ±5 dB at 2ft, at a
frequency of 2500 Hz on
alarm or failure conditions,
may be acknowledged and
then reset by operator
using ACK / RESET button
Output
Sealed 9-pin D connector
provides potential-free
relay contacts for pump
active (pins 6 &7) and
alarm active (pins 8 & 9),
contacts rated for 350V
and 120 mA. Also has
communication lines to
connect to computer for
calibration on pins 1-4.
Dimensions
Instrument: 7.9 x 6.3 x 11.8
in. (20 x 16 x 30 cm) (H x
W x L)
Case: 7.9 x 15.5 x 19.9 in.
(20 x 40 x 51 cm) (H x W x
L)
Weight
Instrument: 9 lb (4 kg) with
attached cables and tubing
Case: 21 lb (9.5 kg) with
detector, 20-ft hose, 35-ft
power cord and manual
Notes:
(
1) Lightly falling rain is defined as ¼ inch/hour for a period of 2 hours at 68 °F (20 °C).
Содержание 3100
Страница 1: ...LUDLUM MODEL 3100 TRITIUM IN AIR MONITOR July 2019 ...
Страница 2: ...LUDLUM MODEL 3100 TRITIUM IN AIR MONITOR July 2019 ...
Страница 11: ...1 2 Figure 1 2 Front View Model 3100 ...
Страница 31: ......
Страница 34: ......
Страница 35: ...5 3 Figure 5 5 Rotary Draw Latches Detail View BF Figure 5 6 Rack Mount Safety Strap ...
Страница 36: ......