POSITION 5
:
AUDIO DIVIDE
-
BY
selects the audible click-per-event division
rate for the current detector setup. If the
AUD ON
/
OFF
switch is in the
OFF
position, no audible click-per-event will be heard.
This parameter ranges from:
0 = Divide By 1
1 = Divide By 10
2 = Divide By 100
3 = Divide By 1000
POSITION 6
:
RESPONSE TIME
allows changing the time constant (TC) for
the current detector setup. If the response is set to 0, the Model 2241-3
automatically calculates (for variable mode) the time constant based on the
incoming cps. If a variable of 1-199 is entered for TC, the response time
becomes fixed.
Variable Response
- Response time is varied in proportion to
the incoming count rate. The two-position
F
/
S
(Fast/Slow)
toggle switch selects the maximum time constant (TC) for the
variable mode. The fast position varies the TC from 4-25
seconds, and the slow position varies from 4-60 seconds.
Fixed Response
- The Fast (
F
) response position is
programmable from 2-50 seconds, and the slow response is 5
times slower than the fast TC. For MDA-type measurements,
the fixed response time mode is recommended.
POSITION 7
:
RATEMETER ALARM
/
ALERT
allows changing the ratemeter
alarm for the current detector setup. The units of this alarm are the same as
the units for the ratemeter display. The fifth push of the left button allows
the decimal point to be moved. The ratemeter alarm adjusts from 1 to 999
R/hr (or Sv/h) or 1 to 999 kcpm or 1 to 100 kcps. The units of the alarm
are determined by the units for the ratemeter.
POSITION 8
:
SCALER ALARM
/
COUNT TIME
sets the scaler alarm variable
from 1-999999, corresponding to the accumulated scaler count. After the
scaler alarm variable is entered, the scaler count time is prompted. The scaler
count time is adjustable from 1-9999 seconds.
POSITION 9
:
NOT USED
Содержание 2241-3
Страница 4: ......
Страница 5: ......
Страница 48: ...Model2241 3 3i TechnicalManual Section8 Ludlum Measurements Inc Page 8 19 January 2016 ...
Страница 49: ...Model2241 3 3i TechnicalManual Section8 Ludlum Measurements Inc Page 8 20 January 2016 ...
Страница 50: ...Model2241 3 3i TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 January 2016 ...
Страница 59: ......
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......