![Ludlum Measurements 2241-3 Скачать руководство пользователя страница 26](http://html.mh-extra.com/html/ludlum-measurements/2241-3/2241-3_manual_1946944026.webp)
Model 2241-3 & 3i
Technical Manual
Section 6
Ludlum Measurements, Inc.
Page 6-2
January 2016
Ludlum Measurements offers a full-service repair and calibration
department. We not only repair and calibrate our own instruments, but most
other manufacturers’ instruments as well.
See Section 8 “Instrument Setup & Calibration,” for further details on
instrument calibration.
Batteries
The batteries should be removed and the battery contacts cleaned of any
corrosion at least every three months. If the instrument has been exposed to
a very dusty or corrosive atmosphere, more frequent battery servicing should
be used. Use a spanner wrench to unscrew the battery contact insulators,
exposing the internal contacts and battery springs. Removing the handle will
facilitate access to these contacts.
Note:
Never store the instrument over 30 days without removing the
batteries. Although this instrument will operate at very high
ambient temperatures, battery seal failure can occur at
temperatures as low as 38 °C (100
°
F).
Содержание 2241-3
Страница 4: ......
Страница 5: ......
Страница 48: ...Model2241 3 3i TechnicalManual Section8 Ludlum Measurements Inc Page 8 19 January 2016 ...
Страница 49: ...Model2241 3 3i TechnicalManual Section8 Ludlum Measurements Inc Page 8 20 January 2016 ...
Страница 50: ...Model2241 3 3i TechnicalManual Section8 Ludlum Measurements Inc Page 8 21 January 2016 ...
Страница 59: ......
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......