Why calibrate?
If the METER is used:
•
With fluids having a viscosity close to the limits of the allowed range
(such as
low-viscosity antifreeze fluids or high-viscosity oils for gear boxes).
•
In extreme flow rate conditions
(close to the min. and max. value of the allowed range), it
may be necessary to carry out an on-site calibration.
How to calibrate
The Meter allows the user to carry out a rapid electronic calibration by modifying the
Calibration factor (K Factor).
At delivery all meters are given the same calibration factor:
K Factor = 1.000
This calibration factor guarantees the best accuracy in the following operating conditions:
Fluid: Motor oil type 10W 30 Temperature: 68°F Flow-rate: 2.6 GPM
The calibration can be done either as:
• An on-site calibration by dispensing into a calibrated container or
• A direct modification of the calibration factor.
On-site calibration by dispensing into a
calibrated container:
Press and hold down six times
to set the function.
The Meter will display:
Blinking point
Press ; start the calibration by
dispensing the fluid into a calibrated
container.
During dispensing the Meter displays:
Batch Total dispensed
Blinking
AUTO
The dispensing operation may be
interrupted and resumed.
The calibrated dispensing is complete
when the level of the fluid reaches the
graduated area of the calibrated container.
The Meter displays:
Indicated value
Blinking
WARNING In order to get a correct
Calibrated K Factor from the
meter, ensure calibration capacity
is set to more than 5 quarts. Purge
all air from the unit before carrying
out the calibration.
Carry out the calibration dispensing at a
steady flow rate by pulling the trigger
completely and keeping it in the open
position until the container is nearly full. Do
not reduce the flow rate to reach the
graduated area of the calibrated container.
Instead, to complete the final stages of the
filling operation into the calibrated
container, top-off in small increments. This
is achieved by rapidly pulling the trigger of
the Meter and then releasing it very quickly.
CRL
Lit
1-000
3.21
Lit
CAL
OK/STOP
AUTO
6.90
Lit
CAL
6.95