[Figure 7-7] 10dB Attenuation Rate Setup
- Example of Use (RBW Selection)
Small signals can be measured when lowering the noise level of the LPT-3000R by reducing RBW.
⒜
Set the frequency of the signal generator to 100MHz and the amplitude to -90dBm.
⒝
Set the Center Frequency to 100MHz.
⒞
Set the Span Frequency to 1MHz.
⒟
Set the Reference level to -30 dBm.
⒠
Execute trace averaging.
⒡
Reduce RBW in the LPT-3000R communication program. The Small signal is more clearly
displayed because the noise level has been reduced.