11.1 Technical data
11.1.1 Device
SPECIFICATION
DEVICE
prisma VENT30,
prisma VENT30-C,
prisma VENT40,
prisma VENT40-C
DEVICE
prisma VENT50
prisma VENT50-C
Product class to MDR (EU) 2017/745
IIa
Dimensions W x H x D in cm
21.8 x 17.5 x 21.8
Weight
2.4 kg
2.5 kg
Weight of internal battery (if present)
0.63 kg
Temperature range
- operation
- Transport and storage
- Transport and storage at +70 °C
- Transport and storage at -25 °C
+5 °C to +40 °C
-25 °C to +70 °C
Allow to cool to room temperature for 4 h
before starting up
Allow to heat to room temperature for 4 h
before starting up
Permitted humidity for operation, transport
and storage
Rel. humidity 10 % to 95 %, no condensation
Air pressure range
600 hPa to 1100 hPa,
corresponds to an altitude of 4,000 m above
MSL (keep leaks small below 700 hPa, as the
device may no longer be able to compensate
at very high ventilation pressures)
Connection diameter for circuit
Standard 22 mm tapered connector to
ISO 5356-1
Maximum air flow at 20 hPa
> 220 l/min
Electrical rating
100-240 V AC, 50-60 Hz,
tolerance
-
20 %
-
10 %
Mean power consumption at maximum
load
At 100 V: 1.02 A
At 240 V: 0.43 A
At 100 V: 1.12 A
At 240 V: 0.5 A
Maximum electrical capacity
100 W
120 W