7 Alarms and faults
Battery
not charging due to
excessive temperature Battery too hot.
Operate device at an ambient
temperature of 5 °C to 40 °C.
Internal battery not
charging - too cold
Battery too cold.
Operate device at an ambient
temperature of 5 °C to 40 °C.
Battery cannot be
charged. Service
necessary
Battery defective.
Have battery replaced.
prismaCONNECT
module defective.
Please get in touch
with your specialist
dealer/contact
prismaCONNECT module
defective.
Have module replaced.
prisma CHECK
module not present. prisma CHECK module
defective or not connected.
Replace module or connect
correctly.
Clock not set.
Internal clock not set.
Have clock set by a specialist dealer
so that course of therapy is recorded
correctly.
Device in battery
mode!
Power supply failed.
Check that the power cord is
securely connected. Check function
of the socket.
Device converted to battery
operation.
Press alarm acknowledgment key.
The device is in battery mode.
Display vanished.
Acoustic and visual
signal for at least
120 seconds, no
display.
Power supply outage and
battery (if present)
discharged.
Check that the power cord is
securely connected. Check function
of the socket. If battery present:
Connect device to power supply and
charge battery.
Device defective.
Have device repaired.
DISPLAY
CAUSE
ACTION