11 Appendix
11.4 Electromagnetic interference immunity for ME equipment
and ME systems
Voltage dips, short
interruptions and
voltage variations in
supply voltage to
IEC 61000-4-11
Number of
voltage drops: 3
drop levels/
duration:
30% / 500 ms
60% / 100 ms
100% / 20 ms
100% / 10 ms at
0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
Number of
voltage drops: 3
drop levels/
duration:
30% / 500 ms
60% / 100 ms
100% / 20 ms
100% / 10 ms at
0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
The quality of the supply
voltage should correspond to
that of a typical business or
hospital environment. If the
user of the device requires
continued FUNCTION, even
in the event of interruptions
to the power supply, it is
recommended that the
device be supplied from an
uninterruptible power supply
or a battery.
Magnetic field at
power frequency
(50/60 Hz) to
IEC 61000-4-8
30 A/m
Duration: 30 s per
axis
Axes: x axis,
y axis, z axis
30 A/m
Duration: 30 s per
axis
Axes: x axis, y axis,
z axis
Magnetic fields at power
supply frequency should
correspond to the values
typical of those found in
business and hospital
environments.
GUIDELINES AND MANUFACTURER DECLARATION - ELECTROMAGNETIC
INTERFERENCE IMMUNITY
The device can be used in both static and mobile operation, in both domestic and
appropriate hospital environments.
In a domestic environment, the device may cause radio interference, possibly making it
necessary to take suitable remedial measures, such as realigning the device, for example.
INTERFERENCE IMMUNITY
TESTS
IEC 60601 TEST LEVEL
COMPLIANCE LEVEL
Conducted HF interference to
IEC 61000-4:-6
10 V
effective value
150 kHz to 80 MHz within
ISM bands
10 V
Radiated HF interference to
IEC 61000-4:-3
10 V/m
80 MHz to 2.7 GHz
80% AM at 2 Hz
10 V/m
GUIDELINES AND MANUFACTURER DECLARATION - ELECTROMAGNETIC
INTERFERENCE IMMUNITY
The device can be used in both static and mobile operation, in both domestic and
appropriate hospital environments.
In a domestic environment, the device may cause radio interference, possibly making it
necessary to take suitable remedial measures, such as realigning the device, for example.