Lowenstein Medical MiniScreen plus Скачать руководство пользователя страница 8

patients. These belts are however also available in
special sizes.

Note:
Overstretching of the belts may result in poor signals
and possible loss of data. 

3.1.6.  Sensor for body position

The position sensor integrated in the MiniScreen unit
displays information related to the patient’s body
position during the study.

For accurate determination of body position, ensure
the MiniScreen is applied to the patient correctly. The
body positions detected are: Supine, Prone, Left
side, Right side, Upright.

3.1.7.  Sensor for leg movement

To diagnose restless or periodic leg movements
(restless leg), the MiniScreen can be equipped with a
leg sensor and internal recording software (optional).
The device then enables continuous recording of leg
movements and associated analysis results in the
report.

Attaching the leg sensor:

The RLS sensor is attached to the leg using the
Velcro wrist band. The sensor should be positioned
approximately 10 cm below the knee joint, lateral to
the shin bone.

Since the leg sensor detects muscle movements,
ensure that the sensor is placed on the muscle and
not the shin bone!

The

red

and

black

safety plugs of the sensor have to

be plugged into the

red

and

black

sockets of the

splitter box.

3.1.8.  Sensor for ECG

The supplied ECG cable (optional) is compatible with
all commonly available adhesive electrodes and can
be used as consumables.

Attaching the ECG electrodes:

The yellow, green and white safety plugs have to be
plugged into the yellow, green and white socket of
the splitter box.

Important:

The ECG is not suitable for a differential

cardiac diagnosis!

3.1.9.  Sensor for Neuroport

The Neuroport signal is obtained via the supplied
EEG cable (optional, MiniScreen plus only). We
recommend special 3-part EEG electrode strips as
consumables.

The 3 part EEG electrode strip needs to be applied to
the forehead. Before applying the electrode the
forehead must be cleaned with a sterile solution. A
headband

around

the

forehead

will

eliminate

movement of the electrode during the study.

The

blue

marked connecting plug of the EEG

electrode cable needs to be plugged into the

blue

socket on the MiniScreen plus unit.

3.2.  Preparing for measurement (ambulatory)

In preparation of an ambulatory study where the
patient will use the device at home, please ensure
the following:

1. Charge the battery: see Page 7.
2. Start the MiniScreen software on the PC.
3. Fill out the fields relating to personal data of the

patient in the menu item ”Record / Initialize
device (Offline)“ and start the transfer.

4. Connect the MiniScreen with the USB interface

cable. For testing purposes, both LEDs will light
up initially. The red LED will then go off. The
green LED on the MiniScreen unit will remain lit
for the duration of the initialization process.

5. A dialogue window will be displayed in order for

the operator to determine the start time for
recording. Once confirmed the unit can then be
disconnected from the PC.

Note

: (remove cable

by pulling on the plug - 

not

 on the cable).

6. The MiniScreen unit is now ready for use. The

unit will now rest in stand-by mode until the
selected Start Time of the recording. The unit
will then automatically switch on and start
recording data. 

3.3.  Operating elements of the device

The MiniScreen device will automatically switch on /
off when connected / disconnected to a PC. The
device has an internal clock for timer recording. If the
timer has been programmed, recording will start
automatically at the predetermined time. Up to eight
recordings can be programmed.

Note: If necessary the MiniScreen can be switched
on by the patient before the selected Start Time by
pressing and holding the Start button for one second.
By pressing the start button (longer than 7 seconds),
the device can be switched off by the patient.

The charging condition of the units battery can be
displayed by pressing the "Akku" button on top of the
device.
Attention: This function will only work when the
device is switched off and disconnected from the PC.

- 6 -

Содержание MiniScreen plus

Страница 1: ......

Страница 2: ......

Страница 3: ...and abdominal movement 5 3 1 6 Sensor for body position 6 3 1 7 Sensor for leg movement 6 3 1 8 Sensor for ECG 6 3 1 9 Sensor for Neuroport 6 3 2 Preparing for measurement ambulatory 6 3 3 Operating...

Страница 4: ...s incl connection cable It is also equipped with the following applied parts SpO2 finger sensor with cable for obtaining pulse frequency and oxygen saturation values Velcro strap for securing the fing...

Страница 5: ...of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper ope ration If such use is necessary this equipment and the other equipment should be...

Страница 6: ...ing the end of its service life the device and its accessories should be disposed of in accordance with the WEEE Directives or relevant Electronic Disposal protocol For further details please contact...

Страница 7: ...turation and pulse frequency A failure of the finger sensor or a missing data update by the pulse oximeter is indicated to the user by the red LED on the device At the same time the values for SpO2 an...

Страница 8: ...be cleaned with a sterile solution A headband around the forehead will eliminate movement of the electrode during the study The blue marked connecting plug of the EEG electrode cable needs to be plug...

Страница 9: ...sensor If applicable attach the abdominal belt leg sensor ECG and EEG electrodes Check that the sensors and their connectors are positioned correctly The next morning The patient has to detach the Min...

Страница 10: ...ominal straps can be washed in the washing machine at 60 C use a laundry net Let the straps dry in the air Attention Do not use a dryer Remove thoracic and abdominal sensor beforehand Disinfection It...

Страница 11: ...tion of the PAP adapter hose on the oxygen mask and on the MiniScreen unit Signal amplitudes are very small or do not appear Check the corresponding sensors on the patient and their connections to the...

Страница 12: ...detection of leg movements restless legs 200 0936 USB cable for data transmission between MiniScreen and PC 100 0815 03 Electrical isolator USB for galvanic separation between MiniScreen device and PC...

Страница 13: ...ensor special rubber coated thimble finger sensor Pulse wave Plethysmogramm display measurement via finger sensor Position Acceleration sensor for position recording 5 positions Ambient Light Photomet...

Страница 14: ...9 Used Symbols 12...

Страница 15: ...SPR 11 Class B The MiniScreen is suitable for the use in all establish ments including domestic establishments and those which are connected to the public power supply which powers buildings for resid...

Страница 16: ...ateur bands 80 AM 1 kHz Radiated RF disturbance IEC 61000 4 3 Proximity fields from RF wireless communications equipment IEC 61000 4 3 10 V m 80MHz 2 7 GHz 80 AM 1 kHz 380 390 MHz 27 V m PM 50 18 Hz 4...

Страница 17: ...0 3 28 5 240 5 100 5 800 WLAN 802 11 a n Pulse Modulation b 217 Hz 0 2 0 3 9 5 500 5 785 NOTE If necessary to achieve immunity test levels the distance between the transmit antenna and the MiniScreen...

Страница 18: ...ance 8 Neuroport 6 Operating elements 6 Optical isolator 3 Ordering informations 10 PAP 2 PAP sensor 5 Position sensor 6 Preparing for measurement 6 Pressure pads 5 RLS sensor 6 Safety related informa...

Страница 19: ......

Страница 20: ...18...

Отзывы: