Lowenstein Medical MiniScreen plus Скачать руководство пользователя страница 11

correct up to date version is selected, open the
software installed and click the Search button. The
program will then automatically find and use the
latest version installed.

After selecting the target drive the installation pro-
gram will now automatically create the necessary
directories and copy all the required files to the hard
drive. The installation will automatically insert a new
group containing an entry for the new software in the
Programs folder of the Start Menu. In addition a
shortcut on your Windows desktop will be created.

After successful installation, remove the CD and
store it in a safe place.

5.2.  Selecting a printer

The measurement curves and reports can be printed
on any commonly available type of printer (ink jet,
dot-matrix or laser printer) that is already installed on
your Windows system. The correct printer driver can
be selected using the Windows Control Panel
application.

5.3.  Viewer Software "Sleep Xpert App"

During an online measurement, the measurement
curves can be displayed on a tablet (Windows or
IOS) optionally with the help of the software "Sleep
Xpert App". The tablet only serves to visualize the
measured curves. There are no data stored and no
findings or evaluation can be made.

6.  Troubleshooting

Channels (e.g. pressure) are missing in the
display.

The channels are deactivated and therefore do not
appear during ”Test” and ”Record”.

Check which channels are active for recording using
the menu item 

Options / Edit Channel sets.

Channels are missing after loading a measure-
ment.

They were not recorded or have been hidden for
measurement data display.

Flow signal is missing during recording or
follows the edge of the range.

The flow signal follows a straight flat line in the
middle of the channel.

Check the flow prong on the patient and the connec-
tion on the MiniScreen unit. When measuring during
PAP therapy, check the connection of the PAP
adapter hose on the oxygen mask and on the
MiniScreen unit.

Signal amplitudes are very small or do not
appear.

Check the corresponding sensors on the patient and
their connections to the MiniScreen unit. Whilst doing
so, check that the tube hoses and the black pressure
pads of the thorax sensors are intact. A leaking sen-
sor can cause inaccurate readings in the thorax
channel. The pressure pads should be attached to

the patient securely without over tightening the belt
and the tube hoses should be routed in such a way
that they cannot kink.

Check all cables and connectors on the MiniScreen
unit and the PC.

Pulse oximeters not responding.

The channels for oxygen saturation and pulse
frequency are registering 50% and 30 P/min respec-
tively despite the patient being connected.

First check that the finger sensor is seated correctly
on the patient and remove any nail polish that may
be present. A small red light should be present in the
sensor when the finger is inserted. If the light doesn't
switch on when the sensor is applied to the finger,
then you need to check the connection at the
MiniScreen unit and any intermediate extensions.

Printout not working.

The printer prints characters on the page but in no
apparent format.

The wrong printer or printer driver has been installed.

The printer is not responding to the print command.

Check the printer cable and the connections to the
printer and PC. The printer should be switched on
and operational, i.e. the control LED’s on the printer
should be lit.

Cannot establish connection to the MiniScreen
unit.

The USB cable is not connected correctly.

Check the connection of the USB cable to the
MiniScreen unit and the PC.

USB interface was deactivated

The

USB

interface

can

be

activated

in

the

MiniScreen device software. Select the "Options" tab
followed by menu "Device Settings"

.

Battery is discharged.

Charge the battery correctly.

For clarification and in case of problems during
the installation, maintenance or use, please refer
to your local distribution partner, or directly to
the manufacturer.

- 9 -

Содержание MiniScreen plus

Страница 1: ......

Страница 2: ......

Страница 3: ...and abdominal movement 5 3 1 6 Sensor for body position 6 3 1 7 Sensor for leg movement 6 3 1 8 Sensor for ECG 6 3 1 9 Sensor for Neuroport 6 3 2 Preparing for measurement ambulatory 6 3 3 Operating...

Страница 4: ...s incl connection cable It is also equipped with the following applied parts SpO2 finger sensor with cable for obtaining pulse frequency and oxygen saturation values Velcro strap for securing the fing...

Страница 5: ...of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper ope ration If such use is necessary this equipment and the other equipment should be...

Страница 6: ...ing the end of its service life the device and its accessories should be disposed of in accordance with the WEEE Directives or relevant Electronic Disposal protocol For further details please contact...

Страница 7: ...turation and pulse frequency A failure of the finger sensor or a missing data update by the pulse oximeter is indicated to the user by the red LED on the device At the same time the values for SpO2 an...

Страница 8: ...be cleaned with a sterile solution A headband around the forehead will eliminate movement of the electrode during the study The blue marked connecting plug of the EEG electrode cable needs to be plug...

Страница 9: ...sensor If applicable attach the abdominal belt leg sensor ECG and EEG electrodes Check that the sensors and their connectors are positioned correctly The next morning The patient has to detach the Min...

Страница 10: ...ominal straps can be washed in the washing machine at 60 C use a laundry net Let the straps dry in the air Attention Do not use a dryer Remove thoracic and abdominal sensor beforehand Disinfection It...

Страница 11: ...tion of the PAP adapter hose on the oxygen mask and on the MiniScreen unit Signal amplitudes are very small or do not appear Check the corresponding sensors on the patient and their connections to the...

Страница 12: ...detection of leg movements restless legs 200 0936 USB cable for data transmission between MiniScreen and PC 100 0815 03 Electrical isolator USB for galvanic separation between MiniScreen device and PC...

Страница 13: ...ensor special rubber coated thimble finger sensor Pulse wave Plethysmogramm display measurement via finger sensor Position Acceleration sensor for position recording 5 positions Ambient Light Photomet...

Страница 14: ...9 Used Symbols 12...

Страница 15: ...SPR 11 Class B The MiniScreen is suitable for the use in all establish ments including domestic establishments and those which are connected to the public power supply which powers buildings for resid...

Страница 16: ...ateur bands 80 AM 1 kHz Radiated RF disturbance IEC 61000 4 3 Proximity fields from RF wireless communications equipment IEC 61000 4 3 10 V m 80MHz 2 7 GHz 80 AM 1 kHz 380 390 MHz 27 V m PM 50 18 Hz 4...

Страница 17: ...0 3 28 5 240 5 100 5 800 WLAN 802 11 a n Pulse Modulation b 217 Hz 0 2 0 3 9 5 500 5 785 NOTE If necessary to achieve immunity test levels the distance between the transmit antenna and the MiniScreen...

Страница 18: ...ance 8 Neuroport 6 Operating elements 6 Optical isolator 3 Ordering informations 10 PAP 2 PAP sensor 5 Position sensor 6 Preparing for measurement 6 Pressure pads 5 RLS sensor 6 Safety related informa...

Страница 19: ......

Страница 20: ...18...

Отзывы: