10.  EMC Information

Medical electrical devices subjected to special preventive measures regarding to electromagnetic compatibility (EMC).
They must be put in operation and installed according to the EMC information in this manual. Portable and mobile RF
communications equipment (e.g. mobile phones) can influence medical electrical devices. Portable RF communications
equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the MiniScreen, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result. For an intended use of the MiniScreen, the specified equipment in this
manual must be used. The application of other equipment can led to a higher emission and a reduction of the immunity. 

Guidelines and Manufacturer‘s Declaration – Electromagnetic Emissions

The MiniScreen is intended for use in an electromagnetic environment as specified below. The customer or user of the 
MiniScreen should ensure that it is used in such an environment. 

Emission Test

Compliance

Electromagnetic Environment - Guidelines

RF –  Emission according to CISPR 11

Group 1

The MiniScreen uses RF energy only for it‘s internal
function. So it‘s RF emission is very low and it‘s
implausible that it cause any interferences to surroun-
ding electronic devices. 

RF –  Emission according to CISPR 11

Class B

The MiniScreen is suitable for the use in all establish-
ments, including domestic establishments and those,
which are connected to the public power supply, which
powers buildings for residential purposes. 

Harmonic Emissions according to 
IEC 61000-3-2

Not applicable

Voltage Fluctuations / Flicker Emissions 
according to IEC 61000-3-3

Not applicable

Table 1: Table 201 EN 60601-1-2, Electromagnetic Emissions

Guidelines and Manufacturer‘s Declaration – Electromagnetic Immunity

The MiniScreen is intended for the electromagnetic environment as specified below. The customer and user of the 
MiniScreen should ensure that it is used in such an environment. 

Immunity Test

Test Specification 

Electromagnetic Environment - Guidance

Electrostatic Discharge (ESD)
IEC 61000-4-2

±8 kV contact
±15 kV air

Floors should consists of wood, concrete or
should be covered with ceramic titles. If the floor
is covered with synthetic material, the relative
humidity must be at least 30 %. 

Electrical fast Transient / Burst
IEC 61000-4-4

± 2 kV for AC power line
± 1kV for input and output line
100 kHz

Surges
IEC 610004-5

± 1 kV differential mode
± 2kV common mode 

Voltage dips
IEC 61000-4-11

0% U

T

 for ½ cycle (1 phase)

0% U

T

 for 1 cycle

70% U

T

 for 25/30 cycles 

a

short interruptions and voltage 
variations on power supply input 
lines
IEC 61000-4-11

0% U

T

 for 250/300 cycles 

a

Power Frequency
(50/60 Hz) Magnetic Field
IEC 61000-4-8

30 A/m

Note

a

U

 is the a.c. mains voltage prior to application of the test level.

25/250 cycles at 50 Hz or  30/300 cycles at 60 Hz

Table 2: Table 4, 5, 7, 8 EN 60601-1-2, Electromagnetic Immunity

- 13 -

Содержание MiniScreen plus

Страница 1: ......

Страница 2: ......

Страница 3: ...and abdominal movement 5 3 1 6 Sensor for body position 6 3 1 7 Sensor for leg movement 6 3 1 8 Sensor for ECG 6 3 1 9 Sensor for Neuroport 6 3 2 Preparing for measurement ambulatory 6 3 3 Operating...

Страница 4: ...s incl connection cable It is also equipped with the following applied parts SpO2 finger sensor with cable for obtaining pulse frequency and oxygen saturation values Velcro strap for securing the fing...

Страница 5: ...of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper ope ration If such use is necessary this equipment and the other equipment should be...

Страница 6: ...ing the end of its service life the device and its accessories should be disposed of in accordance with the WEEE Directives or relevant Electronic Disposal protocol For further details please contact...

Страница 7: ...turation and pulse frequency A failure of the finger sensor or a missing data update by the pulse oximeter is indicated to the user by the red LED on the device At the same time the values for SpO2 an...

Страница 8: ...be cleaned with a sterile solution A headband around the forehead will eliminate movement of the electrode during the study The blue marked connecting plug of the EEG electrode cable needs to be plug...

Страница 9: ...sensor If applicable attach the abdominal belt leg sensor ECG and EEG electrodes Check that the sensors and their connectors are positioned correctly The next morning The patient has to detach the Min...

Страница 10: ...ominal straps can be washed in the washing machine at 60 C use a laundry net Let the straps dry in the air Attention Do not use a dryer Remove thoracic and abdominal sensor beforehand Disinfection It...

Страница 11: ...tion of the PAP adapter hose on the oxygen mask and on the MiniScreen unit Signal amplitudes are very small or do not appear Check the corresponding sensors on the patient and their connections to the...

Страница 12: ...detection of leg movements restless legs 200 0936 USB cable for data transmission between MiniScreen and PC 100 0815 03 Electrical isolator USB for galvanic separation between MiniScreen device and PC...

Страница 13: ...ensor special rubber coated thimble finger sensor Pulse wave Plethysmogramm display measurement via finger sensor Position Acceleration sensor for position recording 5 positions Ambient Light Photomet...

Страница 14: ...9 Used Symbols 12...

Страница 15: ...SPR 11 Class B The MiniScreen is suitable for the use in all establish ments including domestic establishments and those which are connected to the public power supply which powers buildings for resid...

Страница 16: ...ateur bands 80 AM 1 kHz Radiated RF disturbance IEC 61000 4 3 Proximity fields from RF wireless communications equipment IEC 61000 4 3 10 V m 80MHz 2 7 GHz 80 AM 1 kHz 380 390 MHz 27 V m PM 50 18 Hz 4...

Страница 17: ...0 3 28 5 240 5 100 5 800 WLAN 802 11 a n Pulse Modulation b 217 Hz 0 2 0 3 9 5 500 5 785 NOTE If necessary to achieve immunity test levels the distance between the transmit antenna and the MiniScreen...

Страница 18: ...ance 8 Neuroport 6 Operating elements 6 Optical isolator 3 Ordering informations 10 PAP 2 PAP sensor 5 Position sensor 6 Preparing for measurement 6 Pressure pads 5 RLS sensor 6 Safety related informa...

Страница 19: ......

Страница 20: ...18...

Отзывы: