Lowenstein Medical MiniScreen plus Скачать руководство пользователя страница 4

1.  Introduction

1.1.  Assigned Purpose:

The MiniScreen plus and MiniScreen premium Sleep
Diagnosis Systems are recording systems for use in
professional health care environment. They serve as
a differentiated pre-diagnosis for the sleep apnea
syndrome and enable continuous recording of up to
40 channels for at least 15 hours without loss of data.
The recording of data on battery power is also
permitted in home care environment.

They record signals for the following physiological
variables:

Flow

Oxygen saturation SpO

Pulse frequency

Pulse wave

Body position

Respiratory and snoring sounds

PAP (Positive Airway Pressure)

Ambient Light

Thorax effort

Abdomen effort

Breathing frequency

Phase shift Thorax Abdomen

optional:

Thermistor

ECG 

Central heart Frequency

PTT (Pulse Transit Time)

Systolic Blood Pressure (Trend)

Leg movement (LEG)

EEG (Neuroport, only MiniScreen plus)

8 x Analogue channel

13 x channels of prismaLINE

1.2.  Contraindications

There are no absolute or relative contraindications to
using the device. In the following cases, the device
must be used under the supervision of qualified
medical personnel:

in patients with acute life-threatening illnesses.

in patients with acute, serious infections.

in mentally confused patients.

in infants and children

1.3.  Accessories

The MiniScreen system as a whole (ME system) 
contains three components:

The microprocessor controlled MiniScreen 
recording device

PC software for measurement data presentation
and PC analysis

Optional: Analogue box for 8 analogue inputs
incl. connection cable

It is also equipped with the following applied parts:

SpO

2

finger sensor with cable for obtaining

pulse frequency and oxygen saturation values

Velcro strap for securing the finger sensor to the
wrist

Flow prongs for obtaining the respiratory signal.

Adapter hose for measurement during PAP
Therapy

Differential pressure adapter for measurement 
under PAP ventilation

Flexible carry belt with integrated pressure
transducers (thorax effort) to fasten the Mini-
Screen unit to the patient

Flexible belt with integrated pressure transducer
to record abdominal effort

Shoulder bag for storing the MiniScreen

Other accessories:

USB Interface cable for data transfer between
the MiniScreen and PC

Battery recharger

Transport case

Optional applied parts:

LEG sensor for detecting leg movements

ECG electrode for recording ECG signals

Neuroport electrode cable

Thermistor for additional breathing detection

Other optional accessories:

Splitter box for connection the Leg and ECG
sensors

USB isolator for online measurement

Viewer "Sleep Xpert App" for tablet.

The data can be displayed and analysed on a
standard commercial PC. The measurement curves
and the analysis results can be printed on all
commonly available printers, including dot-matrix,
laser or ink jet printers.

The MSV (MiniScreen Viewer) analysis program has
the following minimum configuration requirements:

PC with Windows operating system

Microsoft Windows 7 or higher

512 MB of RAM  (2 GB recommended)

1 GB of disk space on the hard drive

CD-Rom drive for installation

Mouse

128 MByte VGA graphics card with a resolution 
of at least 1024 x 768 (512 MB, 1280 x 1024 / 
True Colour recommended)

Free USB port 

Printer with Windows driver

- 2 -

Содержание MiniScreen plus

Страница 1: ......

Страница 2: ......

Страница 3: ...and abdominal movement 5 3 1 6 Sensor for body position 6 3 1 7 Sensor for leg movement 6 3 1 8 Sensor for ECG 6 3 1 9 Sensor for Neuroport 6 3 2 Preparing for measurement ambulatory 6 3 3 Operating...

Страница 4: ...s incl connection cable It is also equipped with the following applied parts SpO2 finger sensor with cable for obtaining pulse frequency and oxygen saturation values Velcro strap for securing the fing...

Страница 5: ...of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper ope ration If such use is necessary this equipment and the other equipment should be...

Страница 6: ...ing the end of its service life the device and its accessories should be disposed of in accordance with the WEEE Directives or relevant Electronic Disposal protocol For further details please contact...

Страница 7: ...turation and pulse frequency A failure of the finger sensor or a missing data update by the pulse oximeter is indicated to the user by the red LED on the device At the same time the values for SpO2 an...

Страница 8: ...be cleaned with a sterile solution A headband around the forehead will eliminate movement of the electrode during the study The blue marked connecting plug of the EEG electrode cable needs to be plug...

Страница 9: ...sensor If applicable attach the abdominal belt leg sensor ECG and EEG electrodes Check that the sensors and their connectors are positioned correctly The next morning The patient has to detach the Min...

Страница 10: ...ominal straps can be washed in the washing machine at 60 C use a laundry net Let the straps dry in the air Attention Do not use a dryer Remove thoracic and abdominal sensor beforehand Disinfection It...

Страница 11: ...tion of the PAP adapter hose on the oxygen mask and on the MiniScreen unit Signal amplitudes are very small or do not appear Check the corresponding sensors on the patient and their connections to the...

Страница 12: ...detection of leg movements restless legs 200 0936 USB cable for data transmission between MiniScreen and PC 100 0815 03 Electrical isolator USB for galvanic separation between MiniScreen device and PC...

Страница 13: ...ensor special rubber coated thimble finger sensor Pulse wave Plethysmogramm display measurement via finger sensor Position Acceleration sensor for position recording 5 positions Ambient Light Photomet...

Страница 14: ...9 Used Symbols 12...

Страница 15: ...SPR 11 Class B The MiniScreen is suitable for the use in all establish ments including domestic establishments and those which are connected to the public power supply which powers buildings for resid...

Страница 16: ...ateur bands 80 AM 1 kHz Radiated RF disturbance IEC 61000 4 3 Proximity fields from RF wireless communications equipment IEC 61000 4 3 10 V m 80MHz 2 7 GHz 80 AM 1 kHz 380 390 MHz 27 V m PM 50 18 Hz 4...

Страница 17: ...0 3 28 5 240 5 100 5 800 WLAN 802 11 a n Pulse Modulation b 217 Hz 0 2 0 3 9 5 500 5 785 NOTE If necessary to achieve immunity test levels the distance between the transmit antenna and the MiniScreen...

Страница 18: ...ance 8 Neuroport 6 Operating elements 6 Optical isolator 3 Ordering informations 10 PAP 2 PAP sensor 5 Position sensor 6 Preparing for measurement 6 Pressure pads 5 RLS sensor 6 Safety related informa...

Страница 19: ......

Страница 20: ...18...

Отзывы: