During Recording
UTREL30-16 User Guide
32
During Recording
During normal operation, the display shows the most recently recorded temperature. This
temperature is updated at the same rate as the logging interval. The current time is also
displayed (in 24 hour format). A tick symbol is shown as long as no alarm event has
occurred. If an alarm event is registered, a cross symbol is shown instead of the tick.
Below the main temperature the display shows the minimum and maximum temperatures
recorded since the beginning of the trip, or since they were last manually reset.
At the bottom of the display you can see an alarm day summary, where any days on which
an alarm was recorded are highlighted.
Following are some sample display screens:
At 1:25 pm:
l
Alarm events were recorded 6, 18, 19 & 27 days ago.
The following symbols show on the display:
o
for the alarm that occurred 6 days ago
o
for the alarms that occurred 18 and 19 days
ago
o
for the alarm that occurred 27 days ago
l
These alarms were cleared by an inspector, as the
display currently shows .
l
The current temperature is over the upper alarm
threshold, as indicated by the upper alarm marker (
).
l
The duration of this temperature excursion has not
yet triggered an alarm.
At 1:45 pm (20 minutes
later)
l
The temperature has remained above the upper
alarm threshold and has now triggered an alarm
event.
l
The alarm symbol is shown to indicate an alarm
event occurred.
l
The day marker for the current day (
) is shown.