Errors during
updating and
calibrating the
instrument
Freezing during calibration
Generally,
re-calibration
takes
several
minutes. However, it may take more time,
depending on the extent of background
adjustment. If the calibration takes more than
10 minutes, reset the system by turning off and
on using the power button located rear side of
the instrument. Please contact service
engineer if the calibration fails repeatedly.
Incorrect USB drive
Use the USB drive supplied with the
instrument. Or, make sure that your USB drive
is compatible with the instrument. The version
of the USB drive must be 2.0. Some types of
USB drives are not detected or compatible with
the instrument.
More than one firmware
version
Delete previous versions from the USB drive
before downloading new firmware.
Incorrectly saved or
damaged firmware
First, make sure that the USB drive works well
and is compatible with the instrument.
Second, download the file again onto the USB
drive. The file should be located in the root
directory.
Third, ensure that the USB drive is inserted
correctly.
Last, try the update again.
If the problem continues, contact your local
supplier.