Logitech Electronics 2000 MKII Series Скачать руководство пользователя страница 6

Switches and Connections

 

RUN/SET Switch

 

This switch is located in the top right hand corner of the front panel. 

When the instrument is in the 

RUN

 mode, it will display the measured signal, the value of which will 

depend whether the instrument is set to display 

RATE

 or 

COUNT.

   Momentarily pressing this switch 

once puts the instrument into the 

SET

 mode and activates the DP, 0–9 ; and the 

+

 (polarity) switches.  

To return to the 

RUN

 mode, simply momentarily press this switch once again. 

FUNCTION (FUNC) Switch

 

This is located directly below the 

RUN/SET

 switch and is used when it is required to observe or set 

the  programmable  parameters.    When  in  the 

SET

  mode,  momentarily  pressing  the 

FUNC

  switch 

steps through the parameters in the following sequence see 

Programming the Instrument

 on page 8: 

If the instrument is in the 

RUN

 mode, a single depression of the 

FUNC

 switch will toggle the display 

between showing the accumulated 

TOTAL

 value, or the current 

RATE

 value (unless there is no input 

signal when displaying 

RATE

, in which case the display will show  

NO SIG

 ). 

DP (Decimal Point) Switch

 

This  is  located  under  the 

FUNC

  switch  and  is  used  to  set  the  position  of  the  decimal  point  as 

required.   A  single  depression  of  the  switch  will  move  the  decimal  point  one  decade.    Holding  the 
switch depressed will cause the point to move in decade steps from right to left at a pre-determined 
rate.  This switch is only active when setting either the 

Factor

, the 

Total: Scaling Value

 or the 

Rate: 

Scaling Value

 parameters. 

0-9 Switch 

Located  under  each  decade,  these  switches  are  used  to  set  the  value  of  all  programmable 
parameters.  Each depression increments the display one digit.  Holding the switch depressed steps 
the display through  

0

  to  

 at a pre-determined rate. 

+/- Switch 

Located  at  the  lower  left  hand  corner  of  the  front  panel,  this  switch  has  no  function  on 

2000CT 

models.

 

2000CT-

MKII

_hb v3.1  01/2020 

 

 

 

 

!

6

Factor

Universal Scaling Factor

Total....Scale

Multiplying factor for displayed value for TOTAL

Disp. DP

Decimal point position in displayed value for TOTAL

Rate....Scale

Multiplying factor for displayed value for RATE

Disp. DP

Decimal point position in displayed value for RATE

ALA L

“Low Alarm” set point

ALA H

“High Alarm” set point

Содержание 2000 MKII Series

Страница 1: ...2000 MKII SERIES Counter Tachometer Model 2000CT User Guide...

Страница 2: ...arameters 8 FACTOR 9 TOTAL Scaling Value 9 TOTAL Decimal Point Position 10 RATE Scaling Value 10 RATE Decimal Point Position 10 LOW Alarm 10 HIGH Alarm 10 Analogue Output Settings if O P module ordere...

Страница 3: ...al Signal Inputs Signal B count up Control Inputs Factor protection external Reset Control inputs Opto isolated requires sink to 0V of 220W maximum Outputs DC Voltage 10V nominal unregulated 80mA max...

Страница 4: ...of the total display and the rate display so that rate may be displayed as units per second units per minute units per hour milli units per second etc and total can be displayed as units milli units...

Страница 5: ...black Connect the brown wire to the terminal marked with the letter L or coloured red The mains supply to this instrument must be protected with a 1 Amp fuse Transformer Connections 2000 MKII SERIES...

Страница 6: ...RATE in which case the display will show NO SIG DP Decimal Point Switch This is located under the FUNC switch and is used to set the position of the decimal point as required A single depression of t...

Страница 7: ...hown on the display is equal to or greater than the value programmed into memory The LOW alarm operates on a value equal to or below the programmed value The alarm outputs automatically reset when the...

Страница 8: ...ion Other parameters that may need to be programmed are vi Low Alarm limit vii High alarm limit And if the analogue output option has been specified viii Reference ix Full Scale In order to program th...

Страница 9: ...ues that have been programmed prior to returning the instrument to the RUN mode Otherwise when the required values have been programmed press the RUN SET button once to switch the instrument to the RU...

Страница 10: ...ace to the left on each depression of the lower right hand button This sets the position of the decimal point in the display when RATE is being displayed see iii above vi Low Alarm ALA L When the calc...

Страница 11: ...from rate display to total display The unit will display TOTAL for one second followed by the current total count Note RATE is displayed using a maximum of six digits TOTAL is displayed using up to th...

Страница 12: ...reset prior to using the instrument with the revised programming to avoid the possibility of an incorrectly displayed value which has been held in memory from use with the previous settings Protection...

Страница 13: ...ams Front Panel Layout Note On instruments with the SINGLE switch set option only the FUNC switch is accessible unless the bezel and front panel are removed Rear panel layout 2000CT MKII_hb v3 1 01 20...

Страница 14: ...Connection Diagrams Typical Applications Variable reluctance Magnetic Pickups Single input operation Dual input operation 2000CT MKII_hb v3 1 01 2020 14...

Страница 15: ...Alarm Output Connections Alarm connections with external power supply 2000CT MKII_hb v3 1 01 2020 15...

Страница 16: ...advisable to use a good quality mains filter mounted as close to the instrument as possible ensuring that the cable between the filter and the instrument is kept separate from any cables carrying hig...

Страница 17: ...he warranty only extends to the original user Copyright Logitech Electronics Limited 1994 to 2017 Document No 2000CT MKII_hb v3 0 July 2017 2000CT MKII_hb v3 1 01 2020 17 Design and Manufacture of Ele...

Отзывы: