ix — Full Scale (

FS

The setting for Full Scale defines the rate value for which the analogue output will be at maximum 
(5V or 20mA depending on option originally ordered). 

When all parameters have been programmed it is possible to step through all settings by repeatedly 
pressing the centre right hand button (

FUNC

).   When all parameters are set to their required values 

then the unit can be switched to normal operating mode by pressing the top right hand button (

RUN/

SET

). 

Operating Mode 

When the instrument is switched from 

SET

 mode to 

RUN

 mode or when power is applied after the 

programmable parameters have been previously set, the display will show 

RATE

 for a period of one 

second followed by the calculated value or 

NO SIG

 if the input signal does not change state during 

the one second gating period.  If the calculated value is too large to be displayed then the display will 
read 

HIGH

If the 

FUNC

 button is pressed the mode will change from rate display to total display.   The unit will 

display 

TOTAL

 for one second followed by the current total count. 

Note: 

RATE

 is displayed using a maximum of six digits. 

 

TOTAL

 is displayed using up to the full seven digits. 

The rate is continuously monitored and alarms will be set if appropriate.  If the analogue output option 
has  been  specified  then  it  will  be  updated  in  response  to  changing  rates,  irrespective  of  whether 

RATE

 or 

TOTAL

 is shown on the display. 

Reset of Total Count 

It is possible to reset the 

TOTAL

 count by either one of two methods. 

1. 

By momentarily connecting 

RESET

 (terminal 6 on the Rear Panel) to 0V.  This can be 

accomplished by a push button switch or by using a key switch which would provide security 
and avoid inadvertent or unauthorised operation. 

2. 

Alternatively, the 

FUNC

 switch can be used to reset the count.  To zero the 

TOTAL 

display using the

 FUNC 

switch, the button must be held down for approximately six seconds 

while the display shows 

TOTAL 

….

 RESET

 ….

 3 

.…

 2 

….

….

 0

When the display shows  

RESET  0

  the switch button can be released and the 

TOTAL

 count 

value will now be displayed as  

0

2000CT-

MKII

_hb v3.1  01/2020 

 

 

 

 

!

11

Содержание 2000 MKII Series

Страница 1: ...2000 MKII SERIES Counter Tachometer Model 2000CT User Guide...

Страница 2: ...arameters 8 FACTOR 9 TOTAL Scaling Value 9 TOTAL Decimal Point Position 10 RATE Scaling Value 10 RATE Decimal Point Position 10 LOW Alarm 10 HIGH Alarm 10 Analogue Output Settings if O P module ordere...

Страница 3: ...al Signal Inputs Signal B count up Control Inputs Factor protection external Reset Control inputs Opto isolated requires sink to 0V of 220W maximum Outputs DC Voltage 10V nominal unregulated 80mA max...

Страница 4: ...of the total display and the rate display so that rate may be displayed as units per second units per minute units per hour milli units per second etc and total can be displayed as units milli units...

Страница 5: ...black Connect the brown wire to the terminal marked with the letter L or coloured red The mains supply to this instrument must be protected with a 1 Amp fuse Transformer Connections 2000 MKII SERIES...

Страница 6: ...RATE in which case the display will show NO SIG DP Decimal Point Switch This is located under the FUNC switch and is used to set the position of the decimal point as required A single depression of t...

Страница 7: ...hown on the display is equal to or greater than the value programmed into memory The LOW alarm operates on a value equal to or below the programmed value The alarm outputs automatically reset when the...

Страница 8: ...ion Other parameters that may need to be programmed are vi Low Alarm limit vii High alarm limit And if the analogue output option has been specified viii Reference ix Full Scale In order to program th...

Страница 9: ...ues that have been programmed prior to returning the instrument to the RUN mode Otherwise when the required values have been programmed press the RUN SET button once to switch the instrument to the RU...

Страница 10: ...ace to the left on each depression of the lower right hand button This sets the position of the decimal point in the display when RATE is being displayed see iii above vi Low Alarm ALA L When the calc...

Страница 11: ...from rate display to total display The unit will display TOTAL for one second followed by the current total count Note RATE is displayed using a maximum of six digits TOTAL is displayed using up to th...

Страница 12: ...reset prior to using the instrument with the revised programming to avoid the possibility of an incorrectly displayed value which has been held in memory from use with the previous settings Protection...

Страница 13: ...ams Front Panel Layout Note On instruments with the SINGLE switch set option only the FUNC switch is accessible unless the bezel and front panel are removed Rear panel layout 2000CT MKII_hb v3 1 01 20...

Страница 14: ...Connection Diagrams Typical Applications Variable reluctance Magnetic Pickups Single input operation Dual input operation 2000CT MKII_hb v3 1 01 2020 14...

Страница 15: ...Alarm Output Connections Alarm connections with external power supply 2000CT MKII_hb v3 1 01 2020 15...

Страница 16: ...advisable to use a good quality mains filter mounted as close to the instrument as possible ensuring that the cable between the filter and the instrument is kept separate from any cables carrying hig...

Страница 17: ...he warranty only extends to the original user Copyright Logitech Electronics Limited 1994 to 2017 Document No 2000CT MKII_hb v3 0 July 2017 2000CT MKII_hb v3 1 01 2020 17 Design and Manufacture of Ele...

Отзывы: