SCPC System • Error messages and error diagnosis
42
CP_SEM_60-03-00.002
5.1.2
Error codes
The following error messages are displayed on the touch screen (both in the normal user interface and
in the Service menu). In the Service menu this is always displayed in the form of an error code,
whereas in the user interface is given partly as symbols (e.g. battery with a line through it) or plain text
messages (e.g. “No COM!”).
Other error codes
: Error codes from
20
to
28
originate from the SCP System and are stored only in the
SCPC System. Information on these can be found in the Service Manual for the SCP System. Error code
30
indicates a software error following successful completion of the T1 test (switch-on self-test). The
cause for this could be a fault in the EEPROM on circuit board
CPU
.
Error code
Description
Corrective Measures
(please observe sequence)
01
RAM defective
– Replace
ZAU
circuit board
02
EPROM defective, checksum incorrect
– Replace EPROM, if the error remains
displayed:
– Replace
ZAU
circuit board
04
EEPROM write error
– Repeat battery test (one occurrence of
this error is tolerable)
– Replace
ZAU
circuit board if necessary
05
ADC (analog/digital converter) faulty
– Replace
ZAU
circuit board
07
Battery faulty, charge/discharge
interrupted (operation possible without
UPS)
– Check batteries/voltage
– Check battery wiring
– Check fuse S 1 (10 A) on
ZNU
circuit
board
– Check relay on
ZNU
circuit board
– Replace
ZAU
/
ZNU
circuit boards if
necessary
08
09
Saved data (battery test) incorrect
Saved capacity (battery test) incorrect
– Repeat battery test (one occurrence of
this error is tolerable)
– Replace
ZAU
circuit board if necessary
Error in T1 test (switch-on self-test)
10
Current sensor faulty (value > 300 mA)
– Replace
ZNU
circuit board; if the error
display persists:
– Replace
ZAU
circuit board
11
Relay on
ZAU
circuit board faulty, battery
charging not possible
– Replace
ZAU
circuit board
12
13
14
Boost voltage 1/ZAU circuit board
Boost voltage 2/ZAU circuit board
Charge indicator faulty
(Symptom: Charge times too long)
– Replace
ZAU
circuit board
15
Discharge circuit faulty
(Symptom: Battery test not possible)
– Check wiring from
backplane
(CON 4)
circuit board to discharge resistance
– Check discharge resistance
– Replace
ZAU
circuit board
17
Saved offset of current sensor (in EEPROM)
incorrect
– Calibrate current sensor
– Replace
ZAU
circuit board
18
Voltage test delivers different results (2
independent test circuits in the SCPC
System)
– Check mains voltage, if stable and over
190 V (for the 230 V variant):
– Replace
ZNU
/
ZAU
circuit boards
19
Charge time too long
– Check batteries and wiring
Содержание 60-03-00
Страница 1: ...SCPC System Service Manual ...
Страница 12: ...SCPC System Safety 12 CP_SEM_60 03 00 002 ...
Страница 28: ...SCPC System System description 28 CP_SEM_60 03 00 002 ...
Страница 40: ...SCPC System User menu and service menu 40 CP_SEM_60 03 00 002 ...
Страница 44: ...SCPC System Error messages and error diagnosis 44 CP_SEM_60 03 00 002 ...
Страница 82: ...SCPC System CP_SEM_60 03 00 002 ...