SCPC System • Replacing components
68
CP_SEM_60-03-00.002
6.9.2
Mains monitoring (ZNU 2002) circuit board
Replacing fuses
fig. 39: Mains monitoring circuit board fuses
!
Only replace the fuses with the fuses of the same type and specifications! Otherwise damage may be
caused to the SCPC System.
Description
Number
SP number
10 A fuse slow-blow (battery)
S 1
93-511-030
Miniature fuse 1 A, slow-blow
S 2
93-510-039