SCPC System • User menu and service menu
CP_SEM_60-03-00.002
35
Status
AC Power
Battery Power
Power Off
Power Failure
Normal main operation
Battery/UPS operation
Error, ZAU cannot perform shut-down operation (ZAU or ZBP
defective)
Error, ZAU defective/UPS defective
ZAU Normal OP
ZAU T1 Test
ZAU Calibration
Normal function of battery monitoring
Switch-on self-test (for mains operation only)
Running current sensor calibration
Trickle Charge
Battery Test
No Battery
Battery Not Charged
Charge From Deep
Fast I Charge
Fast U Charge
Discharge
Deep Discharge
Maintenance charge (keeps the charging state at 100 %)
Running battery test
No batteries found/batteries faulty
Battery voltage less than approx. 12 V
1st Charge phase, battery voltage less than 18 V
2nd charge phase
3rd charge phase
Discharge, battery/UPS operation
Deep charge activated
Info
Stored Offset [mA]
Stored offset of the current sensor following calibration
(nominal value M ± 70 mA)
Stored Capacity [Ah]
Capacity determined at last battery test
Low Capacity
Capacity determined at battery test less than 80% of nominal
capacity (7.0 Ah)
Discharge Curve Use (in use/possible/not used) Use of recorded discharge curves to determine the calculated
values (e.g. remaining battery time)
Overload Condition
Current overload in battery operation
Possible consequence: Faulty current sensor
Number of Deep Discharges
Saved number of deep discharges, deleted after battery test
Charge Time too long
Charge time over 140% of the normal value
Actual Error Number
Number of current error (see Chap. 5.1.2)
Test in xx days, last: dd.mm.yy
Information on the battery test
Actual Values
Battery Current [A]
Battery current
positive (+) during discharge
negative (–) during charge
Battery Voltage [V]
Current battery voltage
Boost Voltage [V]
Internally generated, increased charge voltage (required for
battery charge for operation on HLM)
Int. Voltage [V]
Internal (currently used) voltage
Cooler Temp [° C]
Temperature on the cooler/charge dropping resistor of battery
monitoring (ZAU)
Power: **/** Change: **
Not relevant for service, internal message
Содержание 60-03-00
Страница 1: ...SCPC System Service Manual ...
Страница 12: ...SCPC System Safety 12 CP_SEM_60 03 00 002 ...
Страница 28: ...SCPC System System description 28 CP_SEM_60 03 00 002 ...
Страница 40: ...SCPC System User menu and service menu 40 CP_SEM_60 03 00 002 ...
Страница 44: ...SCPC System Error messages and error diagnosis 44 CP_SEM_60 03 00 002 ...
Страница 82: ...SCPC System CP_SEM_60 03 00 002 ...