In addition to the numerical result of the pressure is also displayed on the scale of the WHO (Fig. 4). 
WHO scale  is the scale of three-color classification derived values   of blood pressure, as recommended 
by the World Health Organization. Scale, being left on display, allows us to estimate the obtained figures 
according to the classification: the pressure is normal, increased, or this is one of degrees of hyperten-
sion.
6.Touch the 

 – the unit turns off. 

To re-measure repeat the steps in this section. 
TO OBTAIN ACCURATE RESULTS NEEDED A BREAK BETWEEN THE MEASUREMENTS IN 
ORDER TO RESTORE BLOOD CIRCULATION. WHY SPEND REPEATED MEASUREMENTS 
NOT EARLIER THAN IN 3 MINUTES.
The result of each measurement (blood pressure and heart rate) is automatically stored in the 
memory of device.
DATA IN MEMORY WILL BE SAVED EVEN IN THE DEVICE WITHOUT BATTERIES. IN ORDER 
TO DELETE SAVED DATA FROM THE MEMORY DEVICE,FOLLOW THE STEPS DESCRIBED 
IN THE “MEMORY FUNCTION”.
If the power is not switched off and the appliance is not used for 3 minutes, then switches off.

AUTOMATIC INFLATING

If the initial pressure level (190 mm Hg.) is insufficient or there is movement of the hand dur-
ing the measurement, the unit will stop measuring and inflate the cuff to the next higher level 
of pumping. The device has 4  fixed inflation level of the cuffs: 190, 230, 270, 300 mm Hg. 
Automatic inflating of the cuff is repeated until the measurement is completed successfully. This is not a malfunction.

FORCED RELEASE OF PRESSURE FROM THE CUFF

If during the measurement you need to relieve quickly the pressure in the cuff - touch 

. The device release quickly the air from the cuff and off. 

Fig.1

Fig.2

Fig.3

Fig.4

Содержание LD11

Страница 1: ...Digital Blood Pressure Monitor Instruction Manual Ci nieniomierz elektroniczny automatyczny do pomiaru ci nienia t tniczego krwi i pulsu Instrukcja obs ugi LD11 ENG POL...

Страница 2: ...2 ENG NAME OF PARTS AND COMPONENTS 1 Electronic unit 2 LCD display 3 Key M Memory 4 Battery compartment cover 5 Key on off power 6 Cuff 7 Case 8 Battery 9 Instruction manual 10 Warranty card...

Страница 3: ...e from 40 to 160 beats per minute PRINCIPLE OF OPERATION The device uses the oscillometric method for measuring blood pressure and pulse rate with technology Fuzzy Algorithm The cuff with electronic u...

Страница 4: ...l and smoking have a great influence For many people even the procedure for measuring the pressure in the clinic causes an increase in performance Therefore blood pressure measured at home are often d...

Страница 5: ...g soft drinks and alcohol 5 The accuracy of measurement of blood pressure depends on how the size of the device match the size of your arm CUFF IS NOT TO BE SMALL OR CONVERSELY IS GREAT 6 Repeated mea...

Страница 6: ...how the degree of discharge We recommend using alkaline batteries SET THE CLOCK In order to access in the mode of the date and time touch the key M holding it down The selected initial option will fla...

Страница 7: ...uring the measurement keep the left hand slightly with your right hand Pressure measurement while lying 1 Lie on your back 2 Place the cuff at heart level using a case or a folded towel Sit in a chair...

Страница 8: ...ROCEDURE FOR THE MEASUREMENT 1 Before measuring make 5 6 deep breaths and relax Do not move talk and do not strain your arm during the measurement 2 Touch the key 3 The display will briefly highlight...

Страница 9: ...automatically stored in the memory of device DATA IN MEMORY WILL BE SAVED EVEN IN THE DEVICE WITHOUT BATTERIES IN ORDER TO DELETE SAVED DATA FROM THE MEMORY DEVICE FOLLOW THE STEPS DESCRIBED IN THE M...

Страница 10: ...fly displays the average value of all readings stored in memory Fig 5 When you touch the key M again the index 1 the number of memory cells displays briefly on the screen then the result will be the l...

Страница 11: ...t rate deeper vascular sclerosis weak pulse wave the correct blood pressure measurement can be difficult Make sure the cuff is wearing properly and repeat the whole procedure of measurement Repeat the...

Страница 12: ...pecialized organizations 9 On the expiry of the period of service should periodically consult specialists specialized repair organizations to check the technical condition of the unit 10 When disposin...

Страница 13: ...measurement observe the silence and tranquility Repeat the measurement not less than in 5 minutes Unable to produce a large number of measurements The use of low quality batteries Use only alkaline ba...

Страница 14: ...opping Centre Singapore 188307 Manufacturer Little Doctor Electronic Nantong Co Ltd No 8 Tongxing Road Economic Technical Development Area Nantong 226010 Jiangsu PEOPLE S REPUBLIC OF CHINA Distributor...

Страница 15: ...from 20 C below zero to 50 C above zero 85 and below Cuff Model Type Size CUFF LD8 pre formed cuff adult circumference of the wrist 12 5 20 5 cm Overall dimensions Size without cuff Mass without pack...

Страница 16: ...ittle Doctor International S Pte Ltd 2006 2015 LITTLE DOCTOR INTERNATIONAL S PTE LTD Yishun Central P O Box 9293 Singapore 917699 Fax 65 62342197 E mail ld singaporemail com 0123 Shanghai Internationa...

Отзывы: