Operating instructions
4point light 10W (V 4.58)
Page 77 of 130
G
M
L
1. Select electrodes with [
] and [
]
2. Set electrode numbers by using numerical keys.
3. Confirm input with [RET]
If in one line each index is 0 (default setting), end of parameter set is reached.
8.6.5 Run/RunGPRS
Under the menu item “Run” or “RunGPRS” data acquisition in the monitoring mode is
started. The navigation thr
ough the menu “Monitoring – Run” and “Monitoring –
RunGPRS” is much the same. This is why they are treated jointly in this chapter.
In
GPRS
mode
data are transmitted automatically
via
G
eneral
P
acket
R
adio
S
ervice (GPRS)
to a FTP server
after each measurement block. Data transfer is integrated in the measurement,
no interaction of the user with the instrument is necessary..
Select a parameter set: the
parameter set may not be empty
, this means instead of
“Empty!” there has to be a comment in the status line.
Run monitoring
Parameter set : 0
Wenner data acquisition
Fig. 8.76: Select parameter set
1. Select a parameter set with [
] and [
]
2. Confirm selection with [RET]
Measurement param.
Min. volt[mV]: 20.00
Max. error[%]: 0.1
Avg:8 Frq:2.08
Fig. 8.77: Edit measurement parameters
Menu item
Function
Min. volt [mV]
Minimum required receiver voltage
Max. error [%
Maximum permitted error value for U0
Avg: 8
Maximum numbers of averages
Frq: 2.08
Measurement frequency
Tab. 8.59: Measurement parameters
– menu items