Operating instructions
4point light 10W (V 4.58)
Page 55 of 130
G
M
L
Key assignment
Reaction
[
] / [
]
Select parameter
[
]
Delete character
[0]
… [9]
Enter numerical values
[RET]
Continue
[ESC]
Cancel
Tab. 8.38: Navigation through the multimapping parameter set
Up to 10 lines (0 ... 9) of electrode combinations are possible. The instrument will stop
automatically at the first line where all parameters are set to 0 (A0 B0 N0 M0 is the
default setting). For the example above this would be the case in line 4.
3. After editing a parameter file, exit this menu with [ESC ]
5. Start
measurement under “Multimapping” “Run”
8.3.2.3
Delete
Under “Delete” the contents of individual parameter sets can be deleted. The electrodes
A/B/M/N are set to 0 (default setting)
8.3.2.4
Output
After the measurement the data have to be transmitted to the PC. The PC is connected to
the instrument via the serial interface. Data transmission is either in ASCII or binary
format. For transmission in binary format the program
Com4Point
is necessary.
Data transmission with Com4Point
Before the data transfer the terminal program has to be started
→
Start Com4point
on the PC
→
Set
Com4point
into receive mode
→
Start data transfer from the instrument to the PC
MULTI-MAPPING OUTPUT
Single File
All Files
Fig. 8.39: Data transfer