Page 102 of 130
Operating instructions
4point light 10W (V 4.58)
Key assignment
Reaction
[1] / [7]
Select frequency/frequencies
[
] / [
]
Select number of averages
[ESC]
Cancel
[RET]
Confirm input, save values
Tab. 8.86: Navigation through offset calibration
1. Select frequency with [1] and [7]
2. Select number of averages with [
] and [
]
3. Save values and start calibration with [RET]
Offset calibration starts:
OFFSET: 4.16 Hz
I: 0 mA Avg 51
U0 +123.1uV 0.04%
U90 - 5.23 uV 0.02%
Fig. 8.118: Display of offset voltages during calibration
Menu item
Function
4.16 Hz
Frequency
I: 0 mA
Current
, 0 mA means transmitter „off“
Avg 51
Number of averages
U0 +123.1
μV
averaged offset value U0
0.04 %
Typical error in %
U90 -5.23
μV
averaged offset value U90
0.02 %
Typical error in %
Tab. 8.87: Offset calibration menu
Key assignment
Reaction
[ESC]
Cancel
[RET]
End, display and save values