20
Connecting Pressure Tubing
The THMS600
-
PS Stage has a Safety Pressure
Release Valve (1). The safety pressure release
valve will automatically vent excess pressure
above the rated stage pressure.
Note:
Do not tamper with the Safety Pressure
Release Valve or adjust it in anyway.
This will
compromise safety and also invalidate the
warranty
The supplied flexible nylon pressure tubing (6mm
outer diameter; 3mm inner diameter) is connected
to the pressurised gas inlet with a push
-
to
-
fit
connection.
Make sure the end 10mm of the tubing is free of
dirt, grease and has no indentations, scuffs or
defects. Cut the end off with a sharp knife if need
be.
Line up the pressure tubing with the pressurised
gas inlet and push firmly to fix the tubing in
place.
Note: to remove the tubing, hold and pull back
the black locking ring toward the body of the
stage and pull the tubing free of the pressurised
gas inlet.
Push
tubing
Pull
tubing
1