11.2
Controls and Indicators (Control Panel)
For picture of the Control Panel see chapter
5 Control Panel Description
.
11.2.1 Button GO
The GO button
activates the SCU control panel.
After pressing the button
, the SCU control panel will remain active for 3 minutes.
To activate control panel:
Press the button
.
Button/Indicator
Function
Pressure Control Buttons
Changing air pressure in mattress.
MODE Button
Selects desired mode:
APT = Alternating Pressure Therapy Mode
MAX = Maximum Inflate Mode
CLP = Constant Low Pressure Mode
Seat Cushion Indicator
Indicates that seat cushion is connected.
Fowler Boost Indicator
Indicates if backrest is raised.
CPR/ Air Connector Open Indicator
Indicates that air pipe is not connected correctly or not at all.
System Error Indicator
Indicates system error.
Call service department authorized by manufacturer.
Mute Button
Muting audio alert signals.
Service Due Indicator
Indicates that service is due.
GO Button
Activates the control panel.
Accumulator Charge Status Indicator
(accumulator back-up model only)
Indicates charge status of accumulator.
Pressure Level Indicators
Indicate the air pressure set in mattress. Pressure Level Indicators are active
only in APT Mode or CLP Mode. Pressure Level Indicators are disabled in
MAX Mode or when Seat Cushion is connected to SCU.