20/46
D9U001SW0-0101_02
8.3 Transporting the Bed
CAUTION!
Damage to the system due to incorrect preparation before moving!
The bed may be moved only after the Ethernet cable has been disconnected from the local computer network (if
it is connected to the local network with cables) and after the bed has been disconnected from the electricity.
When moving the bed (e.g., from one room to another, from room to theatre, etc.), it is essential to disconnect the bed
from the local computer network (if cable connection is used) and the network cable from the electricity. When the bed is
being moved, the given bed is not monitored or localised.
8.2 Tag Installation
CAUTION!
The Tag box must be installed upright (1). The distance between the bottom edge of Tag and the fl oor must be
17 cm!
If the distance from the fl oor is smaller, the proper function of the system cannot be guaranteed and the bed can
be displayed in a position other than the one it is actually in!
CAUTION!
Every Tag has its serial number!
Make a note of the serial numbers of the installed Tags along with their positions! Tag serial number is listed on
the back on the Tag ID.
Tag installation steps:
►
Specify the Tag position on the wall. Tag must be placed on the wall directly in the middle of the bed's parking
position. The distance between two Tags must be at least 2.2 meters. The Tag bottom edge must be just 17 centimetres
above the fl oor.
►
Remove Tag from the box (
1
).
►
Glue Tag to the correct position on the wall. If it is necessary to adjust the wall so Tag holds reliably, use dowels
and screws (
3
) to attach a tin plate to the correct position on the wall (
2
), to which Tag is then glued.
1
3
2
Fig. Bottom of the Tag box (serial number)
Fig. Contents of the Tag box