D9U001SFS-0101_03
87
12 Cleaning/Disinfection
WARNING!
Risk of injury due to accidental bed movement!
►
Always disable the function buttons of a medical bed when cleaning SafeSense® 3 system components (Hub
Box, Air Pad, Wetness Pad and cables) installed on the medical bed.
CAUTION!
Material damage due to incorrect cleaning/disinfection!
►
Do not use washing machines.
►
Do not use pressure or steam cleaners.
►
Follow the instructions and observe the dosages recommended by the manufacturer.
►
Ensure that disinfectants are selected and applied exclusively by qualified hygiene experts.
►
Check if used cleaning agents and disinfectants are compatible with materials that the product consists
of! For information see the following table.
CAUTION!
Risk of material damage due to cleaning of the SafeSense® 3 system in a washing machine!
►
Before machine washing of a medical bed, always remove the components of the SafeSense® 3 system (Hub
Box, Air Pad, Wetness Pad and cables) from the medical bed on which the system is installed!
CAUTION!
Risk of material damage due to steam cleaning of the SafeSense 3 Monitoring System!
►
Always remove the components of the SafeSense® 3 Monitoring System (Hub Box, Air Pad, Wetness Pad and
cables) from the Eleganza 2 medical bed on which the system has been installed!
CAUTION!
Risk of material damage due to incorrect washing of the Wetness Pad!
►
It is recommended to wash the Wetness Pad at temperature of 60°C.
►
Do not tumble dry for maximum lifetime of the Wetness Pad (50 washes). Washing of the Wetness Pad at tempe
-
rature of 85°C with medium heat tumble dry causes shorter lifetime of the Wetness Pad (30 washes).
►
Do not use fabric softener. Do not make the Wetness Pad waterproof.
►
Hang Wetness Pad to dry after washing.
PRODUCT COMPONENTS THAT
ARE INTENDED TO BE CLEANED
Do not clean what is not mentioned
in this column!
MATERIALS (SURFACES OF THE MENTIONED COMPONENTS)
Competent user is responsible for check if used cleaning agents and disinfectants
are compatible with mentioned materials!
Air Pad
Polyester (PES)
Air Hose
Polyvinylchlorid (PVC)
Wetness Pad
Cotton
Hub box
Acrylonitrile butadiene styrene (ABS)
Hub holder
Lacquered steel + Polyamide (PA)
Cables (LAN cable, Nurse Call cable,
Wetness Pad cable)
Polyvinyl chloride (PVC)