D9U001SFS-0101_03
44
ORDER OF THE CONNECTION OF THE FOUR WETNESS CABLE CORES
Order
1
2
3
4
Cable core
marked by this
label
To connect Wetness Pad to the SafeSense® 3 system:
►
Connect the RJ connector on the opposite side of the Wetness Pad cable to the Hub (connector in the position
3
).
►
Connect 4 connectors of the Wetness Pad cable to the four connectors on the Wetness Pad according to labels on the
cable.
9.3 Eleganza 2 with SafeSense® 3
WARNING!
Risk of injury and risk of material damage due to incorrectly installed Hub box!
►
SafeSense® 3 Hub must be placed on the Hub holder and these two components must be fixed on the accessory
rail on the right bed side under the seat section!
►
Avoid collisions between Hub holder and accessories of the bed and between Hub box and accessories of the
bed!
►
Ensure the indispensable cables are correctly and safely connected to the Hub box!
WARNING!
Risk of injury and risk of material damage due to incorrect installation of cables and incorrect installation of
the Air Hose!
►
Ensure the cables leading to the SafeSense® 3 Hub are not kinked or sheared or otherwise machanically da
-
maged during bed positioning or during manipulation with accessories of the bed!
►
When routing cables from SafeSense® 3 system in the Eleganza 2 medical bed avoid squeezing those cables
between parts of the medical bed!
►
Cables of the SafeSense® 3 system and Air Hose must not be tensioned!
►
Ensure the Air Hose is correctly connected to the Hub box and to the Air Pad!
►
Avoid squeezing, kinking, shearing of the Air Hose or other mechanical damage to the Air Hose!
WARNING!
Risk of injury and risk of material damage due to incorrect installation of Air Pad and Wetness Pad!
►
Use only mattress compatible with SafeSense® 3 system!
►
Air Pad must be fixed on the Backrest of the Eleganza 2 medical bed. Prevent the Air Pad from moving on the
Backrest under the mattress!
►
Ensure the Air Hose is correctly connected to the Air Pad fixed on the Backrest of the Eleganza 2 medical bed!
►
Wetness Pad must be placed on the seat section of the compatible mattress and the ends of the Wetness Pad
must be put under the seat section of the compatible mattress.
►
Wetness Pad must not be covered with a plastic foil!
►
Ensure the cable connecting Wetness Pad with the SafeSense® 3 Hub is correctly connected to the SafeSense®
3 Hub and to the Wetness Pad.
CAUTION!
Risk of material damage due to steam cleaning of the SafeSense 3 Monitoring System!
►
Always remove the components of the SafeSense® 3 Monitoring System (Hub Box, Air Pad, Wetness Pad and
cables) from the Eleganza 2 medical bed on which the system has been installed!
Содержание SafeSense 3
Страница 41: ...D9U001SFS 0101_03 41 Fig Air Hose connected to the Hub Fig Wetness Pad cable connected to the Hub...
Страница 52: ...D9U001SFS 0101_03 52 Fig Air Hose connected to the Hub Fig Wetness Pad cable connected to the Hub...
Страница 81: ...D9U001SFS 0101_03 81 Fig License section of the System Settings screen...