LightWave EasyCheck Fiber Optic End-face Inspector
LightWave EasyCheck Fiber Optic End-face Inspector
Image Capture
CAP is the image capture key. Pressing CAP can capture and save the current image to the
SD card. The screenshot can only be accessed with the SD card inserted. Power must be
switched off before insertion and removal of the SD card.
Inspections
After adjustment, connect the product to be tested onto the adaptor tip and adjust the focus
ring to get the clear image. Spin the X-axis screw clockwise, connector moves towards the
left; spin the X-axis screw counterclockwise, connector moves towards the right; spin the
Y-axis screw clockwise, connector moves upwards; spin the Y-axis screw counterclockwise,
connector moves downwards; users can view the entire surface condition with the help of XY
axis screws.If polishing is required, press CAP during inspection and the image captured will
be kept on file in SD card. Power off the inspector, pull out SD card and reinsert into
computer.
Picture View
Press to enter the picture view mode. Images stored in SD card can be viewed in this
mode. Press MENU to exit.
CAP:
Delete: pressed the first time, DEL? prompt is displayed; Press again to delete the
current image.
:
Up key, View the previous image.
:
Down key, View the next image.
External Monitor Instructions
In this mode, images can be viewed simultaneously from the external monitor and the
EasyCheck’s 8" display. Connect the external monitor to EasyCheck’s AV OUT port tip with AV
video cable. Turn on the EasyCheck and the external monitor. Connect the simplex or multi-
channel optical fiber connector with the EasyCheck and adjust the focusing ring to get a clear
image. The end-face status can then be viewed from both displays. Make sure the image
display switch is turned to the right in this mode.
External Optical Fiber End-face Inspector Instructions
Easycheck’s has two AV IN port types: USB and AV. The USB type is customized for external
end-face inspectors. Connect external end-face inspector and EasyCheck and make sure the
EasyCheck’s image display switch is turned to the left. Power on the Easycheck and connect
the fiber connector to the external end-face inspector. Adjust the Easyget's focus ring to get
the clear image so as to observe the end-face condition.
RCA version has general image input tip for all types of end-face inspectors. Connect
inspector with EasyCheck and make sure the image display switch is turned to the left. Power
on the Easycheck and connect the fiber connector to the general inspector. Adjust the focus
ring to get the clear image so as to observe the endface condition.
Chapter 4 | Maintenance
Notice
1. Instruments must be moved gently.
2. No extrusion.
3. Power must be switched off when taking out SD card.
4. Use the Dimension Technology customized 12V DC power adapter. Other power
adapters are not allowed.
5. Do NOT disassemble instrument so as to avoid unnecessary loss.
6. Keep instrument away from sharps so as to avoid scratches on the screen.
Maintenance
1. Keep Easy check clean, away from grease and dust.
2. It would be better kept in the instrument container if the inspector will not be used for
a long time.