TruScan 532 Single - Wavelength – Operators Manual Rev. No 02
Page 39 of 114
between each pulse of the treatment laser. Similar to the Duration switches, the
left hand control will decrease the interval and the right hand control will increase
the interval. In this mode the repeat pulse rate cannot be set higher than a 50% duty
cycle. This means that the laser fire duration cannot be higher than the repeat interval
duration. It means also that the repeat interval duration is greater than or equal to the
exposure duration when pressing the down setting switch. The minimum interval is
0.01 seconds, and the maximum interval is 3.0 seconds. If the interval is adjusted
beyond 3.0 seconds, then the Repeat Interval display will show “One” where the
footswitch must be pressed every
time the doctor/user wishes to fire the laser.
9. Aiming Beam Intensity Indicator display
This display gives an indication of the brightness of the Aiming beam intensity. When
more of the bar graphs are displayed, then the brightness is higher. At the maximum
brightness the output power will be around 1.0mw or less to the patient’s cornea.
10. Aiming Beam Intensity button
The brightness of the Aiming beam is adjusted with these 2 switches.
11. LIO Illumination Intensity Bar Indicator
This bar graph is the intensity indicator.
12. LIO Illumination Setting button
When the LIO is connected to the TruScan 532 Single-Wavelength Laser Console,
this control can be used to adjust the intensity of the Illumination lamp.
13
.
Warning Volume Control Button / Display Bar
This control can be used to adjust the volume of the laser fire audible warning.
However the audible warning cannot be turned off.
14. Total Laser Counts display
This display gives a sum of the total number of Laser shots fired during the laser
treatment.
15. Counter Reset Button
Whenever this switch is pressed the total laser shot counter is reset to zero.
16. Type of Delivery System Indicator
This indicator will indicate the type is being used.
Содержание TruScan 532
Страница 1: ...Operator s Manual TruScan 532 Single Wavelength ...
Страница 74: ...TruScan 532 Single Wavelength Operators Manual Rev No 02 Page 73 of 114 ...
Страница 80: ...TruScan 532 Single Wavelength Operators Manual Rev No 02 Page 79 of 114 Figure 5 15 b System Ready to operate ...
Страница 115: ...TruScan 532 Single Wavelength Operators Manual Rev No 02 Page 114 of 114 ...