TruScan 532 Single - Wavelength – Operators Manual Rev. No 02
Page 106 of 114
8.2 Warning
A completed listing of Lightlas system warming and the suggested corrective action are
detailed as following:
A complete listing of the TruScan 532 Single-Wavelength system warning along with the
suggested corrective actions are shown below:
Error code
Cause(s) / Description
Remedy Action(s)
Safety filter not
found
-Safety filter not connect
to the console or loosen
with console
-Please connect the safety filter to the
console.
Red diode
temperature not
ready
-Diode is not according to
ambient temperature
-Please wait while the console adjusts
itself according to ambient
temperature
-This process may take up to five
minutes to complete.
LBO temperature
not ready
-LBO is not according to
ambient temperature
-Please wait while the console adjusts
itself according to ambient
temperature
-This process may take up to five
System
Warning
Message
Corrected
Actions