TruScan 532 Single - Wavelength – Operators Manual Rev. No 02
Page 92 of 114
Installation Record Sheet
Model: TruScan 532 Wavelength
Console Serial Nunber: __________________
LIO Serial Number: _________________
Slit lamp LDU Serial Number: ______________
1.
System Setup
(Tick for OK)
1.1 No damage to external box packaging
1.2 All parts or module received.
1.3 Environment and electrical
power validation
2
.
Table Top & Columns
2.1 Unopened all the accessories and module
2.2 Validation if all the received and well-conditioned
2.3 Assembled all the appropriate modules
2.4
Plug in the power & validate if up / down and power switch LEDs lit
2.5
Slit is fully adjustable to closed position
2.6 Reset the columns by pressing down key to the end and columns
integrity
2.7 Elevate the column to the max height & lower again to validate the
table
3.
Slit lamp Check
3.1 Illumination lamp control checkout
3.2 Filters control checkout
3.3 Appecture settings control checkout
3.4 Slit centering & opening and closing checkout
3.5 Slit rotation ch/- 90 degrees
3.6 Slit lamp movement checks (left to right or vice versus)
4. Chinrest Check
4.1 Chinrest fully secured with screws
4.2 Chinrest height is adjustable & movement smoothness
4.3 Fixation
lamp integrity & movement
4.4
Arm handlers are fully secured
5. Remote / console checks (optional)
5.1
Plug in the remore control into console
5.2
Power on the console and validate if the display its integrity
Содержание TruScan 532
Страница 1: ...Operator s Manual TruScan 532 Single Wavelength ...
Страница 74: ...TruScan 532 Single Wavelength Operators Manual Rev No 02 Page 73 of 114 ...
Страница 80: ...TruScan 532 Single Wavelength Operators Manual Rev No 02 Page 79 of 114 Figure 5 15 b System Ready to operate ...
Страница 115: ...TruScan 532 Single Wavelength Operators Manual Rev No 02 Page 114 of 114 ...