Set up the Counter
248083387-1 Rev 10
4-23
15. Type
N
in response to “Change telnet config password” as shown
in Figure 4-23. The instrument is shipped without a password.
Changing the telnet password or providing one is discouraged,
except for absolute security requirements. If a password is
applied and forgotten, the instrument will have to be returned to
Lighthouse to get the password cleared.
Figure 4-23 Telnet Config Password Screen
16. Type
S
to Save and restart and press
Enter
to save the changes.
See Figure 4-24.
Note:
When the
Ethernet restarts, it will
lose connection to the PC,
reported as, “Connection
to host lost”.
Figure 4-24 Saving the Settings
17. Perform Step 7. though Step 16. for each instrument.
WARNING:
Step 18 is
very important to prevent
network address errors
on the PC when the
process is complete.
18. Clear the ARP Cache by type
arp –d *
and press
Enter
.
If this is
not done before quitting this process, the PC may create “ghost
IPs” on the LAN and cause serious problems with the network.
19. Type
Exit
to quit the command console.
Содержание SOLAIR 3350
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions SOLAIR 3350 5350 Gen E Airborne Particle Counters Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Страница 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Страница 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Страница 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Страница 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Страница 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Страница 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Страница 183: ...00...
Страница 184: ......