Lighthouse SOLAIR 3350, 5350 Gen E Operating Manual
7-6
248083387-1 Rev 10
The low word register for Sample Time is 40034.
The low word register for Hold Time is 40032.
The low word register for Initial Delay is 40030.
Table 7-4 Instrument Parameters
Register
Data Type
Description
40026
unsigned integer
Location number; Specifies location of
Particle Counter.
40029
unsigned integer
Initial Delay [high]. Works in conjunction
with 40030. Number of seconds to wait
before starting the first sample. Max value
is 86,399, which equals 23h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction
with 40032. Number of seconds to wait
between sample periods. Max value is
86,399, which equals 23h 59m 59s
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in
conjunction with 40034. Number of
seconds to sample. Max value is 86,399,
which equals 23h 59m 59s.
40034
unsigned integer
Sample Time [low]
Содержание SOLAIR 3350
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions SOLAIR 3350 5350 Gen E Airborne Particle Counters Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Страница 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Страница 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Страница 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Страница 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Страница 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Страница 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Страница 183: ...00...
Страница 184: ......