Set up the Counter
248083387-1 Rev 10
4-9
EU GMP 2009
Note:
The EU GMP
2009 report does not apply
to the Boulder Counter - it
does not support a particle
size of 0.5
μ
m.
The SOLAIR supports the EU GMP 2009. This standard applies only
to 0.5
μ
and 5.0
μ
channel sizes.
If either one of these channel sizes is not present, the report cannot
classify the room. These requirements preclude running this report on
the Boulder Counter - its size channels start at 5.0
μ
.
There are only 4 Grades: A, B, C, D, and 2 cleanroom states: At Rest
and In Operation.
The standard defines a minimum sample size of 1m
3
for the Grade A
cleanroom and refers to ISO standards for the rest of the grades. In
order to take a measurement, at least 1 location and 1 sample are
required by the counter. The volume of air must be shown in liters.
The classification for a particle size is determined based on the mean of
all average concentrations taken from all locations in the cleanroom. If
the mean exceeds all limits, then the classification is “Out of Range”.
The counter will only print channel sizes of 0.5 and 5.0 in this report.
Ethernet
Configuration
The SOLAIR comes pre-configured to use
DHCP and receive an IP address
automatically. If this is not a desired
configuration, use this section to change
these settings.
If these terms are not understood nor the
possible impact of changing the SOLAIR’s
settings, do NOT proceed without
contacting the facility’s IT personnel.
The following procedure requires pre-
approval and planning by the IT
department and a joint setup effort.
Содержание SOLAIR 3350
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions SOLAIR 3350 5350 Gen E Airborne Particle Counters Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual t vi 248083387 1 Rev 10...
Страница 14: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual ii 248083387 1 Rev 10...
Страница 22: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 2 6 248083387 1 Rev 10...
Страница 64: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 5 6 248083387 1 Rev 10...
Страница 142: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual 6 78 248083387 1 Rev 10...
Страница 154: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual A 2 248083387 1 Rev 10...
Страница 176: ...Lighthouse SOLAIR 3350 5350 Gen E Operating Manual D 2 248083387 1 Rev 10...
Страница 183: ...00...
Страница 184: ......