Lighthouse REMOTE Liquid Particle Counter (MODBUS Output) Operating Manual
7-8
248083227-1 Rev 8
Note:
The REMOTE
comes standard with 2
particle channels; the table
below shows the optional 4
channels.
The threshold registers (45xxx series) run in parallel with the data
registers (30xxx series). For example, data register 30010’s
corresponding threshold register would be 45010. Data register 30016’s
threshold register would be 45016. See Table 7-8.
Setting the Alarm Threshold Value
The Alarm Threshold Value is set in the low register of the channels.
Multiple channels can be enabled for alarms. Different thresholds can
be set for each channel. See Table 7-9.
Table 7-8 Alarm Threshold Registers
Register Data
Type
Description
45009
unsigned int Threshold for Particle Channel 1 [high]
(smallest particle size starts here)
45010
unsigned int Threshold for Particle Channel 1 [low]
45011
unsigned int Threshold for Particle Channel 2 [high]
45012
unsigned int Threshold for Particle Channel 2 [low]
45013
unsigned int Threshold for Particle Channel 3 [high]
45014
unsigned int Threshold for Particle Channel 3 [low]
45015
unsigned int Threshold for Particle Channel 4 [high]
45016
unsigned int Threshold for Particle Channel 4 [low]
Table 7-9 Alarm Threshold Registers set to 1000
Registers
Particle
Channel
Threshold
Value
45009 - 45010
1
1000
45011 - 45012
2
1000
45013 - 45014
3
1000
45015 - 45016
4
1000
Содержание RLPC 0.1m
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counter MODBUS Output Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual LOF 2 248083227 1 Rev 8...
Страница 14: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual ii 248083227 1 Rev 8...
Страница 24: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 2 8 248083227 1 Rev 8...
Страница 32: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual 3 8 248083227 1 Rev 8...
Страница 112: ...Lighthouse REMOTE Liquid Particle Counter MODBUS Output Operating Manual D 2 248083227 1 Rev 8...
Страница 117: ...00...
Страница 118: ......