Getting Started
248083317-1 Rev 4
3-9
18. The hub end of the cable run can now be terminated. After both
ends are terminated, the wire should be checked using an ethernet
100baseT cable tester (not just a continuity tester) to ensure that the
signals and ground will be reliable. Steps 6 through 12 can be used
as a reference for termination of the hub end. Figure 3-13 illustrates
how the wires should be connected.
Figure 3-13 RJ45 to Cat5 Pinout
19. Make sure the hub power is OFF or the hub end of the cable is
disconnected from the equipment before connecting the cable to the
instrument, but do NOT connect it, yet. Connecting cables is
discussed later in this chapter.
Mount the
REMOTE-4PN
The REMOTE-4PN Series uses a NEMA style enclosure and mounting
to a flat vertical surface is required for proper operation. Pump flow
control is managed by the REMOTE-4PN Series pump controller.
Power to the REMOTE-4PN Series is managed by a power switch on
the NEMA enclosure bottom plate. The cover lock uses a flat blade or
coin to turn the latch to the open or closed position. Do NOT apply
power to the REMOTE-4PN Series until cables are attached properly at
both ends.
Sample tubing runs should terminate at the instrument location and
should use Bevaline tubing to prevent static and particle dropout. The
REMOTE-4PN Series instrument supports remote monitoring when
equipped with a tubing kit of 3 meters.
Содержание Remote 3014PN
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions REMOTE 4 PN Series Airborne Particle Counter Operating Manual...
Страница 6: ......
Страница 10: ...Lighthouse REMOTE 4PN Series Operating Manual t iv 248083317 1 Rev 4...
Страница 12: ...Lighthouse REMOTE 4PN Series Operating Manual ii 248083317 1 Rev 4...
Страница 58: ...Lighthouse REMOTE 4PN Series Operating Manual 4 20 248083317 1 Rev 4...
Страница 66: ...Lighthouse REMOTE 4PN Series Operating Manual A 2 248083317 1 Rev 4...
Страница 78: ...Lighthouse REMOTE 4PN Series Operating Manual B 12 248083317 1 Rev 4...
Страница 82: ...Lighthouse REMOTE 4PN Series Operating Manual I 4 248083317 1 Rev 4...
Страница 83: ...00...
Страница 84: ......