ApexRp Operators Manual
61
248083480-1 Rev 3
Additional status bits are shown in 40056 displayed in Table A-6.
Bit
Description
0
ApexRp:
LASER POWER STATUS
: Set to 1 when unit’s LASER
current is out of spec, else set to 0
1
ApexRp:
LASER CURRENT STATUS:
Set to 1 when unit’s
LASER power is out of spec, else set to 0
2
ApexRp:
LASER SUPPLY STATUS:
Set to 1 when unit’s LASER
supply is out of spec, else set to 0
3
ApexRp:
LASER LIFE STATUS:
Set to 1 when unit’s LASER
supply is out of spec, else set to 0
4
ApexRp:
NO FLOW STATUS:
Set to 1 when unit’s flow is below
no flow threshold causing LASER to turn off, else set to 0.
5
ApexRp:
PHOTOAMP SUPPLY STATUS:
Set to 1 when unit’s
photo amp supply is out of spec, else set to
6
ApexRp:
BACKGROUND STATUS:
Set to 1 when unit’s photo
amp background is out of spec, else set to 0
7
ApexRp:
PHOTODIODE STATUS:
Set to 1 when photodiode has
failed, else set to 0
8
ApexRp:
CALIBRATION DUE DATE STATUS:
Set to 1 when
unit is past calibration due date, else set to 0
9
ApexRp:
LOCATION BRACKET STATUS:
Set to 1 when unit is
in location bracket mode and bracket is missing, else set to 0.
Table A-6 Device Status (40056)
Содержание ApexR02p
Страница 1: ...Operators Manual ApexRemoteWithPump A I R B O R N E P A R T I C L E C O U N T E R W I T H P U M P...
Страница 2: ......
Страница 10: ...ApexRp Operators Manual 9 248083480 1 Rev 3 Intentionally Blank Page...
Страница 14: ...ApexRp Operators Manual 13 248083480 1 Rev 3 Intentionally Blank Page...
Страница 22: ...ApexRp Operators Manual 21 248083480 1 Rev 3 Intentionally Blank Page...
Страница 38: ...ApexRp Operators Manual 37 248083480 1 Rev 3 Intentionally Blank Page...
Страница 73: ...ApexRp Operators Manual 248083480 1 Rev 3 72 Intentionally Blank Page...