16
Data for electromagnetic compatibility (EMC)
EMV-Allgem eine Daten
Emission
Übereinstimmung
RF-Emission according
EN 55011
group 1
RF-Emission according
EN 55011
Class B
Generation of mains harmonics
follow ing IEC 61000-3-2
Battery operation, not applicable
Generation of mains harmonics
follow ing IEC 61000-3-3
Battery operation, not applicable
EMV-Tabelle 202
Susceptibility
IEC 60601-test level
actual level
ESD
IEC 61000-4-2
±6kV cd
±8kV ad
±2kV,±4kV ±6kV cd
±8kV ad
Bursts
IEC 61000-4-4
±2kV mains
±1kV I/O
Battery operation, not applicable
Surges
IEC 61000-4-5
±1kV dm
±2kV cm
Battery operation, not applicable
Voltage drops etc.
IEC 61000-4-11
Reduction to
Reduction to
5% for 10ms/ positive
amplitude
Battery operation, not applicable
5% for 10ms/ negative
amplitude
Battery operation, not applicable
40% for 100ms
Battery operation, not applicable
30% for 500ms
Battery operation, not applicable
0% for 5000ms
Battery operation, not applicable
H-field at 50/60Hz
IEC 61000-4-8
3A/m
no m agnetically sesitive sensors
or parts, not applicable
EMV-Tabelle 204
Susceptibility
IEC 60601-test level
Actual level
Conducted RF
IEC 61000-4-6
3Veff
150kHz to 80 MHz
3Veff
Radiated RF
IEC 61000-4-3
3V/mf
80 MHz to 2,5 GHz
3V/m
EMV-Tabelle 206
Output pow er of
transmitter
W
Safety distance depending on frequency/m
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2,5 GHz
0,01
0,12m
0,12m
0,24m
0,1
0,37m
0,37m
0,74m
1
1,17m
1,17m
2,34m
10
3,69m
3,69m
7,38m
100
11,67m
11,67m