
T4 EEG Amplifier User Manual
24
Figure 10 Setup Recording dialog
The order of the signal labels in the pull down list is the same as the order of the signals in the sig-
nal list defined using the signal-editing tool.
Amplifier setup: amplifier setup activates the recording inputs in preparation for a recording. For
most applications you need perform amplifier setup only once – when the system is first installed –
since the amplifier setup is saved with the recording montage for future recall and usage. See Step
4 below.
If you want to use the recording channel order defined in amplifier setup, click on Auto Record
Setup in the Setup Recording dialog box. You can then skip Step 4 (Define the recording channels).
Poly and AUX inputs: these inputs can be set to either referential (EEG mode), bipolar AC or bi-
polar DC. They are ideal for polygraphic signals such as respiration, airflow, EKG, body position (DC
mode) etc.
To calibrate an AUX input to reflect a required unit of scale for a given voltage input use the signal
editor – see Step 2). Select the appropriate units, e.g. %, or mm Hg, and enter the Physical Signal
Amplitude required to generate the Amplifier Input Voltage.
Also in the Aux input section is the SaO
2
channel for the Nonin Xpod unit. Select this channel if us-
ing the Xpod. Note: If a Setup that uses greater than 32 channels plus the Nonin is attempted to
be downloaded to the recorder, an error is displayed and the download prevented.
Step 4 Define the recording channels
Step 4 can usually be skipped, since the Auto Record Setup button will copy what you have defined
under the inputs in Step 3 into the list of recording channels.
However, you can define and save recording montages for spe-
cific recording needs, and recall them for future usage.
Creating a montage follows the same principle as the signal
creation and input definition tool: click on the channel number
to define the active and reference label of choice.
An example of a recording montage is shown below.
Figure 11 Channel setup