T4 EEG Amplifier User Manual
20
3.5
Shutdown of the system
At the completion of a study proceed as follows to shut down the system:
•
Stop the recording by pressing the ‘PC Record Off’ button
•
Exit the Trackit program.
•
Shut down Windows.
•
Switch off the PC and disconnect the mains supply.
After the recording is finished, the Micro-SD memory card is removed for reading in a PC.
3.6
Battery replacement and charging
For ambulatory, body-worn applications the T4 is powered from a USB Power Bank. The T4 plugs
directly into the USB port on the Power Bank. Ensure that this uses the 1A socket on the Power
Bank. After connection, press the green ‘ON’ pushbutton on the side of the battery.
1
2
Open zip at bottom of bag to access battery
Insert charged battery into bag
Unplug cable from battery
Plug cable into battery
Withdraw discharged battery from bag
Turn battery on using button on side
Do not connect cable to any other equipment
Close zip
WARNING: The Amplifier must only be used with the USB Power Bank supplied or au-
thorised by Lifelines.
•
Do not short circuit this device. To avoid short circuit, keep the device away from any metal
objects (e.g., hair clips and keys).
•
This device may get hot during use, and this is normal. Hold carefully.
•
This unit is not user repairable.
•
Do not heat this device or throw it into a fire.
•
Do not drop or place the unit under a heavy object.
•
Keep this device away from high temperature, wet, or dusty environments.
•
During charging keep the device out in the open to allow excess heat to dissipate.
•
Charge M5 for 4 hours / M10 for 8 hours before initial use.
•
Keep the USB output port and micro-USB input port clean and free of obstruction.
•
For additional instructions, consult the documentation supplied with the battery.
To replace the battery in the Bag, unzip the bottom of the Bag to gain access to the battery con-
nection. Unplug the USB connection to the battery and replace with a fully charged unit.
When not connected to the T4 unit, the battery is charged by using the micro-USB socket and con-
necting to a standard USB port, using the supplied cable.
Battery Charging
Connect the micro-USB charging cable supplied to the Battery and use a USB connection for charg-
ing. The battery capacity display will illuminate to show how much of the battery is charged. Once
all the LEDs are lit, the battery is fully charged and is ready for use.
For additional instructions, consult the documentation supplied with the battery.
Do not charge the battery when used in the Home Environment.
WARNING: Do not charge the battery whilst it is connected to the T4 unit.
WARNING: Do not charge the battery when the equipment is in use.