38
DM00003 Rev D
SYSTEM
MANUAL
11�6 Photic Stimulator
Category
Specification
Power Requirements
y
Voltage
12 Vdc
y
Current
2 A
Recommended Distance from Patient
20 cm
Stimulation Light Intensity
y
LVL 1 Setting (Nominal)
3,000 lux
y
LVL 10 Setting (Nominal)
10,000 lux
Stimulation Frequency
y
Minimum
1 Hz
y
Maximum
60 Hz
y
Resolution
1 Hz
Stimulation Pulse Width
y
Minimum
0.1 ms
y
Maximum
12.8 ms
y
Resolution
0.1 ms
Stimulation Duration (Per Frequency)
y
Minimum
1 second
y
Maximum
16 seconds
y
Resolution
1 second
Rest Duration (Between Frequencies)
y
Minimum
1 second
y
Maximum
16 seconds
y
Resolution
1 second
Technical Description