4�2 Cautions for the Trackit M
Proper usage of this system depends on careful reading of all instructions and labels.
A power interruption that occurs during an EEG recording session may cause non recoverable
data loss. If you experience frequent power interruptions, Lifelines Neuro recommends usage of an
Uninterruptible Power Supply (UPS). To minimize the risk of data loss, follow all the requirements of the
UPS operation manual including the timely replacement of the batteries.
Do not autoclave or sterilize, except for aseptic techniques on cabling and electrodes, any parts of
Trackit M.
Do not use acetone or other solvents for cleaning the components of the system. Cleaning the system
without first turning it off may result in electric shock or system malfunction.
Installation of additional software on the unit other than those explicitly approved for use on the device
may result in system malfunction. Do not delete or modify any system files as this may also result in
system malfunction.
Do not run other applications while conducting a study with the Rendr Acquisition Software or the
system may not function correctly.
Data collected during electrostatic discharge to the system may contain spike noise artifact.
Data collected within environments containing high frequency noise may be subject to such noise. In
such cases, the traces may appear thicker.
Always disconnect all cables, turn off the device, and close the screen before transportation of the
Trackit M to avoid injury and system damage. Also, to avoid system damage, only transport the system
using approved packaging.
Adjusting the date and time of the system before conducting a study will affect the date and time
information associated with the study. Ensure that the date and time of the system is correct before
beginning a study. Do not change the date or time while performing a study.
Do not enable the screen saver or turn off the system while performing a study. Only turn off the system
according to the procedure outlined in this manual.
Do not expose the system to large magnetic fields or it may result in lost data or system malfunction.
Do not block the fan vents on the back and rear of the device or it may overheat, leading to system
malfunction.
The manufacturer does not recommend the use of extension cords when powering the Trackit M or its
accessories, as it may result in decreased noise performance or an electrical safety hazard.