Section 5. Troubleshooting
Code 128 triggered
Code 128 is triggered if the inlet is obstructed. This can happen if the instrument
has drawn water, dust, or debris into the inlet filter, causing a clog. In either case,
power off the instrument, clear the obstruction, replace the filters (see
on page 6-4), and power the instrument back on.
You must restart the instrument to resolve code 128.
5-6
LI-7820 N₂O/H₂O Gas Analyzer