Section 3.
Applications of the LI-7820
In this chapter we provide information about using the instrument in field or labor-
atory applications. We cover connecting plumbing to the air inlet, retrieving data,
and other applications.
Flow schematic
During normal operation, air is drawn into the analyzer through the air inlet (
below). Air flows through the optical bench and phase adjuster and is
exhausted through the air outlet. In the optical bench, the pressure is drawn down
to about 40 kPa. When the instrument is powering down, air is circulated in a
closed loop through the desiccant to ensure that air remaining in the system is free
of moisture, preventing condensation when the instrument is powered off.
Figure 3-1. Simplified flow schematic of the gas analyzer. During a measurement, (
A
) air
flows through the optical bench and phase adjuster. During shutdown (
B
), the instrument
purges the optical path of water vapor.
3-1