Section 6. Maintenance
4
Remove the column.
It is secured with two screws. Remove them. When the column mounting brackets
are removed, the column can come apart easily, and the silica gel beads can fall
out. Be careful when handling the column so that it does not come apart.
5
HOLDING THE COLUMN AWAY FROM THE CASE, separate one end by grasping one cap and
pulling it off of the column.
Pour out the silica gel beads. They can be disposed of in municipal waste.
6
Locate a new packet of silica gel beads in the accessories kit.
7
Fill the column with fresh silica gel beads.
There may be a little more than is needed. Dispose of the extra silica gel beads.
8
Reassemble the filter and caps; connect the tubes, and tighten the screw clamps securely.
Important:
Tighten the clamp screws securely. The clamps hold the caps onto
the tube; the caps hold the silica gel beads in the tube. If the clamps are not
tight, the caps will be forced off when the tube is pressurized during shutdown.
The tube will open and the silica gel beads will spill in the case. It is a situation
easily avoided by tightening the clamp screws securely.
6-6
LI-7820 N₂O/H₂O Gas Analyzer
Содержание LI-7820
Страница 1: ...LI 7820 N2O H2O Trace Gas Analyzer Instruction Manual ...
Страница 2: ......
Страница 10: ...viii ...
Страница 28: ...Section 2 Getting started 2 12 LI 7820 N O H O Gas Analyzer ...
Страница 36: ...Section 3 Applications of the LI 7820 3 8 LI 7820 N O H O Gas Analyzer ...
Страница 52: ...Section 4 Software overview 4 16 LI 7820 N O H O Gas Analyzer ...
Страница 66: ...Section 6 Maintenance 6 8 LI 7820 N O H O Gas Analyzer ...
Страница 84: ...Appendix B Additional safety information B 6 LI 7820 N O H O Gas Analyzer ...
Страница 91: ......