Coin-cell battery
The product includes a CR Lithium coin-cell battery, which contains perchlorate
material. Special handling may apply. In the U.S. State of California, see
www.dtsc.ca.gov/hazardouswaste/perchlorate.
To remove the coin-cell batteries, power off the instrument, remove the batteries,
and open the lid to the case. Remove the four screws that secure the electronics
shroud. Being careful to avoid straining the Wi-Fi antenna (it is glued to the
shroud), move the shroud to the side. The battery is near the middle of the instru-
ment. The battery is held in with friction. Press it out. Insert a replacement battery
with the positive (+) pole facing up.
Figure B-3. Carefully lift the shroud as shown to avoid straining the Wi-Fi antenna, which
is attached to the shroud.
Appendix B. Additional safety information
B-5